OVERVIEW 
 

Below are the steps you need to follow to begin using your Kensington KS-EC-10L-E pocket conductivity tester. This 
manual will cover how insert the batteries, using the function keys, making measurements, calibration and sensor 
replacement.  
 

INSERTING THE BATTERIES 

 
 
 
 
 
 
 
 
1. Your KS-EC-10L-E pocket pH tester comes with one G13a button battery. Remove it from the package. 
2. Remove the button cap found on the bottom of the tester as shown in the above image. 
3. Place the battery into its designated slot. Take note of the proper position of the battery to ensure the polarities 
    placed correctly. 
4. Replace the battery cap. 
 

KEYPAD INDEX 
 
 
 

  
 
 
 
 
 
 
 
 
 
 

Key 

Function 

 

1. Serves as the power button of the tester 
2. Initiates Data Hold. Press once to hold the current reading, press again to resume making 
measurements. 
3. Exit button during calibration 

 

1. Press once to conduct calibration of the instrument 
2. Press the button again to choose the type of buffer used in the calibration. 

 

1. Press to show the current temperature during measurement mode 
2. Decreases the setting value in calibration mode. 

 
 
 
 
 
 
 

Содержание KS-EC-10L-E

Страница 1: ... Remove it from the package 2 Remove the button cap found on the bottom of the tester as shown in the above image 3 Place the battery into its designated slot Take note of the proper position of the battery to ensure the polarities placed correctly 4 Replace the battery cap KEYPAD INDEX Key Function 1 Serves as the power button of the tester 2 Initiates Data Hold Press once to hold the current rea...

Страница 2: ...on 146 5 µS cm is required stir the electrode gently 3 Make sure there are no air bubbles trapped in the slot of the sensor 4 Press and hold the CAL key The tester will now enter into calibration mode 5 Press or to set the calibration value 6 Ensure the displayed value matches your calibration standard Press MEAS to confirm 7 The calibration value will then automatically flash three 3 times and th...

Страница 3: ... storage solution when the tester is not in use 2 If the instrument will be kept in storage for a long time place the sensor in an electrode storage solution 3 Make sure to wash the sensor with distilled water before and after use 4 Do not reuse prepared buffer solutions ELECTRODE REPLACEMENT One of the advantages of using the Kensington KS EC 10L E is that the sensor can be replaced instead of bu...

Страница 4: ... E Range 1 0 199 9µS cm Accuracy 0 1 FS Calibration points 1 point Calibration solutions 146 5µS cm Power off Manual or Automatic 8 minutes after last key pressed Temperature compensation 0 50 C Temperature coefficient 2 C Cell constant K 1 Normalization Temperature 25 C Hold function Manual Operating temperature 0 60 C 32 140 F Power requirements 3 1 5V G13A Batteries Dimensions 185 L 40 Dia mm W...

Отзывы: