General Cleaning Directions
Cleaning of KANGAROO ePump enteral feeding pumps must be performed as follows:
CAUTION: Disconnect pump from AC power source before cleaning. After cleaning, do not connect to
AC power source until pump and power cord are thoroughly dry.
• A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and
hypochlorite mixture, however, repeated cleaning with this solution can damage the plastic pump case. Isopropyl
alcohol, applied with a damp (not wet) cotton swab, may be used for cleaning difficult-to-reach areas, however, it
should be used sparingly because repeated cleaning may damage the case.
• DO NOT USE strong cleaners such as Spray-Nine™*, Phisohex™*, Hibiclens™*, or Vesta-Syde™* because damage to
the pump case housing can result.
Directions for Cleaning Pump Housing
• Refer to General Cleaning Directions before starting.
• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as
possible.
• Avoid excess moisture near pole clamp area.
• Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.
Directions for Cleaning Pump Power Cord
• Refer to General Cleaning Directions before starting.
• Unless soiling is observed, the power cord should not be cleaned.
• If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with a
cloth dampened with isopropyl alcohol.
CAUTION: Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or
fire hazard.
Directions for Cleaning Rotor Assembly
• Refer to General Cleaning Directions before starting.
• Open the blue door enclosing the Pump Set loading area.
• Loosen rotor set screw with 5/64” (2mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid
getting any moisture in the rotor shaft opening.
• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
• Be sure all parts of rotor are completely dry before putting it back onto shaft.
• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
• Push rotor into place and tighten set screw (Do not over tighten).