INSTALLATION INSTRUCTIONS

MRIPSF-480_F-7499_031919

IMPORTANT SAFEGUARDS

When using electrical equipment, basic safety precautions should always be followed, including the following:

 

THIS PRODUCT MUST BE INSTALLED IN ACCORDANCE WITH THE APPLICABLE INSTALLATION CODE BY A PERSON FAMILIAR WITH 

THE CONSTRUCTION AND OPERATION OF THE PRODUCT AND THE HAZARDS INVOLVED. DISCONNECT POWER TO ALL CIRCUITS 

BEFORE WIRING FIXTURE. INSTALL IN ACCORDANCE WITH ALL NATIONAL, STATE, AND LOCAL CODES. DO NOT CONNECT TO 

AN UNGROUNDED SUPPLY. READ ALL FIXTURE MARKINGS AND LABELS TO ENSURE CORRECT INSTALLATION OF FIXTURE. 

SUPPLEMENTAL INSTRUCTIONS MAY BE LOCATED ON THE FIXTURE, IN ADDITION TO THIS INSTRUCTION SHEET, REGARDING 

ORIENTATION, OR MOUNTING RESTRICTIONS. 

 

SAVE THESE INSTRUCTIONS

INSTALLATION INSTRUCTIONS

MEDMASTER

MRIPSF SERIES

Remote Power Supply and Filter System for MRI Luminaires

This instruction sheet applies to the Kenall MRIPSF-480 Remote Power Supply & EMI Filter System. It describes the 
proper mounting and wiring procedure for a typical MRI room application. A separate instruction sheet is provided 
with the MRI luminaire(s). Both sheets together are intended to provide comprehensive instructions on system 
installation and electrical wiring. 
The remote supply & EMI filter must be installed outside the shielded MRI environment. It contains ferrous 
components that can damage or interfere with MRI equipment. Complete conduit runs and wiring between 
power supply, EMI filters and luminaires after all system components have been installed in their final positions. 
If luminaires are intended to be dimmed (via 0-10V dimming system), then the Kenall MRIFD-1A EMI filter (or 
equivalent) must be ordered and installed.

 .875 KO (9)

 1.125 KO (7)

 2.93

2.05

2.00

2.00

 8.62

END VIEW

BACK VIEW

SIDE VIEW

.88

15.18

4.05

1.05

12.00

1.60

1.29

8.10

.31 Dia. Mounting Holes (4 places)

6.00

2.67

3.25

3.00

3.25