7.3.4
Checking the compressed air safety valve
The compressed air vessel with the compressed air safety valve is located in
the rear area of the product.
Fig. 33: Access to the compressed air compressed air safety valve
Pos. Description
Pos. Description
1
Maintenance cover
3
Knurled screw
2
Compressed air safety valve
Tab. 26: Positions on the product
To gain access to the compressed air vessel/safety valve (Pos. 2), please
proceed as follows.
1. Remove the right maintenance cover (Pos. 1) of the product using a
suitable tool.
2. Loosen the knurled screw (Pos. 3) on the compressed air safety valve
(Pos. 2) by turning it anti-clockwise and open it by approx. 3 - 4 turns
until the exhaust air process starts. (audible discharge of compressed
air)
3. Allow the compressed air safety valve to blow off briefly.
4. Screw in the knurled screw (Pos. 3) as far as it will go and hand-tighten
it.
5. Close the maintenance cover (Pos. 1) and put the product back into
operation. See chapter “Commissioning”