Operating Instructions for Helical Flow Meters Type SRZ
KEM
9
5 Service and Maintenance
5.1 Maintenance
The SRZ* does not require regular maintenance.
For best performance, however, we recommend a recalibration every year if operated 24 hours a day
or every 2 years if the unit is running less than 24 hours a day.
When used with aggressive or abrasive media the SRZ* should be checked and recalibrated even
more often.
If there is no flow over a longer period, the SRZ* should be emptied and cleaned. This is especially
important if curing media like paints are measured.
For recalibration and test, please contact your nearest dealer or directly KEM.
For the addresses see chapter 6.6
5.2 Trouble
shooting
In case the VTE* does not work properly, first check the following items:
No output signal
All cables properly connected?
Î
connect the missing cables
VTE* properly mounted?
Î
Check mounting position and, if necessary, mount the VTE* properly
Output frequency too low
All cables properly connected?
Î
reconnect the loose cables
Meter working properly?
Î
Check and, if necessary, replace the meter
Output frequency too high or unstable
Most probably EMC problems
Shield and ground properly connected?
Î
Connect shield properly. If necessary, try additional means of grounding and shielding
5.3 Service
The SRZ* does not contain any user serviceable parts.
In case of malfunction, please contact your nearest dealer or directly KEM.
For the addresses see chapter 6.6