5.  Menu navigation for calibrators 

 

6. Commissioning

Switch the device on 
Press the SELECT button to switch the 

device on. Initially, the device shows 

the pressure range calibrated in the 

factory (top) and the software version 

(year / week).

Switch the device off 
Keep the SELECT button pressed 

down until the display shows 

OFF

Press the ENTER button to execute 

the shutdown.

 The settings made previously are 

retained when you switch the device 

on and off.

Display mode 
Display mode is the calibrator’s basic 

mode. The upper part of the display 

shows the pressure unit and the pres-

sure that is currently measured. The 

lower part of the display shows the last 

function that was used, either the min./ 

max. display or the Leak function. 

Using the functions 
Written descriptions of the individual 

functions are given below (in addition 

to the diagram above).  

Selecting functions 
The individual sub-functions are called 

up from the MANO menu. Keep the 

SELECT button pressed until 

MANO

 is 

ENTER button 
The ENTER button positioned on 

the front is used to activate the se-

lected function or pressure unit on the  

device. You can also press the ENTER 

button to switch between the mini-

mum and maximum pressure values  

measured thus far.

16

Menu navigation for calibrators

 Pressure unit

 Current measured value

S E L E C T   b u t t o n

E

N

T

E

R

 

b

u

t

t

o

n

❶  

Resetting peak- and trough pressure

❷  

Min. / max. display

❸  

Leak measurement

❹  

Set zero point permanently

❺  

Reset zero point to factory setting

❻  

Activate / deactivate switch-off function   

❼  

Select units

 Leak runtime in mm:ss

 Pressure change within 

leak runtime

Содержание MPX

Страница 1: ...Operating instructions for the medium pressure calibrator MPX...

Страница 2: ...3 Operating the MPX calibrator 15 4 Description of the functions 15 5 Menu navigation for calibrators 16 6 Commissioning 16 7 Maintenance disposal 17 8 Software for calibrators 18 Technical data 19 Sp...

Страница 3: ...e cable K 104A or K 114A PC connection RS485 2 Device base 3 Display 4 SELECT and ENTER buttons 5 Hand pump integrated in the device 6 Vacuum pump 7 Pressure relief valve 8 Fine tuning valve 9 Pressur...

Страница 4: ...t Zero To re set the pressure zero point to absolute pressure use the RES Zero function reset zero Commissioning A pressure resistant connection for the test object is required in order to use the med...

Страница 5: ...pres sure connection for the test object 9 Connect the test object and the vacuum pump 6 to the Y coupler Use the vacuum pump to lower the pressure You can then fine tune the pressure by screwing the...

Страница 6: ...n max display or the Leak function Using the functions Written descriptions of the individual functions are given below in addition to the diagram above Selecting functions The individual sub function...

Страница 7: ...are Software for calibrators MANO Continuous function Automatic switch off function the device switches off automatically 15 minutes after a button was last pressed Leak measurements are canceled by t...

Страница 8: ...delivery K 114A or can be downloaded free of charge at www keller druck com Settings on the medium pressure calibrator with the ManoConfig software Device settings such as the leak measurement time or...

Страница 9: ...e Fischer cable socket on the side fits the K 104A K 114A interface converter Compensated temperature range 0 50 C Operating temperature 0 50 C Storage temperature 10 60 C Air humidity 5 95 relative h...

Страница 10: ...x Test object adapter G 1 4 M G 3 8 F 506810 0028 x Test object adapter G 1 4 M G 1 2 F 506810 0013 x Sealing ring G 1 8 508635 0001 x x Sealing ring G 1 4 508635 0002 x x CrNi filter 307025 2011 x x...

Отзывы: