LEX 1 has two operating keys. The left key

(SELECT)

serves to select

the functions and the pressure units. The right key 

(ENTER)

activates

the selected function or pressure unit. The right key is also used to
switch between the Max.- and Min.-value.

Turn-on:

Pressing the SELECT key turns the instrument on. The

instrument first displays the software version (year/week) and the full-
scale pressure range. The instrument is then ready for use and indi-
cates the actual pressure (top display) and the last measured Max.-
value (bottom display).

The instrument has the following functions:

RESET:

Max.-/Min.-value are set to the actual pressure

OFF:

Turns off the instrument

MANO:

Releases the following functions:

ZERO SEt:

Sets a new Zero reference

ZERO rES:

Sets the Zero to factory setting

CONT on:

Deactivates the automatic turn-off function

CONT off:

Activates the automatic turn-off function

(the instrument

turns off 15 minutes after the last key function)

followed by the unit selection:

bar, mbar, hPa, kPa, MPa, PSI,

kp/cm

2

, cmH

2

O, mH

2

O, inH

2

O, ftH

2

O, mmHg, inHg

Example: Setting a new pressure unit (mbar):

-->  Turn on the instrument by shortly pressing the SELECT-key.
-->

Wait for the instrument’s measuring mode (approx. 3 seconds).

-->  Press the SELECT-key 3 times:

MANO

appears.

--> Press 

ENTER:

ZERO SEt

appears.

--> Press 

SELECT:

ZERO rES

appears.

--> Press 

SELECT:

CONT on

or 

CONT off

appears.

--> Press 

SELECT:

bar

appears.

--> Press 

SELECT:

mbar

appears.

-->  Press ENTER: The new pressure unit (mbar) is set.

The instrument returns to the measuring mode.