Qu
] -> ATD function [
Q.Atd
]
Parameter
Function
Default
[
T.DEL
]
Time delay [s]
1 s
(
T.ACT
)
Sampling time [s]
15 s
► Press [Mode] to access the main menu
> The display shows [
Ao
]
► Press the [▼] until [
Qu
] is displayed
► Press [Mode]
> The display shows [
Q.EPS
]
► Press [▼] until [
Q.ATD
] is displayed
► Press [Mode]
> The display shows [
t.dEl
]
► Press [▼] until [
t.Act
] is displayed
► Press [Mode]
> The current sampling time is displayed
► Press [▲ or ▼] for 2 second
> The display flashes three times
► Press [▲ or ▼] to set the desired sampling time
► Press [Mode] to confirm the parameter
Exit operation
► Wait 30 seconds
or
► Press [▲ or ▼] to change to the parameter [
End
]. Then Press [Mode] to change to the
functional menu.
► In the functional menu press [▲ or ▼] to change to the parameter [
End
], then press
[Mode].
Alternatively, the parameters can be set via IO-Link.
10 Adaptation of temperature level
By changing the emissivity ratio, the difference between the measured temperature level and
true temperature can be compensated.
For this purpose, a comparison measurement with an immersion lance is required. Adjust
emissivity ratio in such a way that the measured values coincide. An increase in the emissivity
ratio results in a low temperature value.
The graph shows an approximation of the influence of the emissivity ratio setting on the change
in temperature value.
Quick Start Guide CellaCast PA 8x
Содержание CellaCast PA8 Series
Страница 1: ...Pyrometer CellaCast PA8x 0000000 12 2021...
Страница 2: ......
Страница 48: ...RU CellaCast PA 8x 1 X 8x CellaCast PA www keller de its 2 3 2 1 0 630 680 4 CellaCast PA 8x...
Страница 49: ...RU CellaCast PA 8x 1 5 30 30 50 6 65 C 1 2 3 4 6 125...
Страница 51: ...RU CellaCast PA 8x 8 PA xx AF xx C PA xx AF xx L 2 2 DIN 916 DIN 911 1...
Страница 55: ...RU 1 2 C 101 104 2 103 0 IO Link CellaCast PA 8x...
Страница 56: ...RU 11 KELLER HCW IO Link CellaCast PA 8x...
Страница 57: ...CN CellaCast PA 8x 1 PA 8x www keller de its 2 3 1 0mW 630 670nm 4 CellaCast PA 8x...
Страница 58: ...CN 1 5 30 30 50 6 65 C 1 2 3 4 6 125 7 24V DC 18 32 V DC CellaCast PA 8x...
Страница 60: ...CN PA xx AF xx L 2 2 DIN 911 DIN 916 1 60 C LED 2 15 9 5 s 1 2 1 T DEL 2 CellaCast PA 8x...
Страница 63: ...CN 30 End End IO Link 10 1 2 C CellaCast PA 8x...
Страница 64: ...CN 101 104 103 0 IO Link 11 KELLER HCW IO Link CellaCast PA 8x...
Страница 66: ...KR 1 5 30 30 6 bar 50 l min 65 C 1 Target size 2 3 field of view 4 6 125mm CellaCast PA 8x...
Страница 68: ...KR CellaCast PA 8x 8 through lens target marker PA xx UF xx C PA xx UF xx L MODE 2 MODE 2 DIN 911 DIN 916 O 1...
Страница 72: ...KR CellaCast PA 8x 1 Ratio correction 2 C 101 104 2 103 0 IO Link 11...
Страница 73: ...KR CellaCast PA 8x KELLER HCW IO Link...
Страница 74: ......
Страница 75: ......