www.keller-druck.com
Version 08/2020
Seite 13 von 24
5
RECORD Data Storage
The record data storage offers the advantage that the measuring data doesn’t get lost if the data transfer (connection
to the gateway) is temporarily out of function, the data can also be read out on site. All measured values are stored in
th
e ADT1’s EEPROM. The memory is organized as a circular memory. This means that always the latest data is available
whilst the oldest data is overwritten.
5.1
Storage Capacity
The table below shows you how much data can be stored in the ADT1 memory.
Storage interval
Number of channels
Number of measured values per channel
Recording time
10 min
4
24552
85 days
10 min
14
8184
28 days
1 h
4
24552
510 days
1 h
14
8184
170 days
5.2
Read the RECORD data by USB
Connect the ADT1 with the USB interface and start the "KOLIBRI Desktop" software. The device should be displayed
under “Generic Ports”
.
The connection to ADT1 is now established automatically and all available RECORDs appear. Select the desired RECORD
and click on the
button “Start reading selected record from device”
.