7 Shielding and Grounding
7.1 Equipotential bonding
The pyrometer housing is connected to the shielding via the cable connector!
Differences in ground potentials might cause an equalising current to flow between
devices through a cable shielded at both ends� In this case, be sure to install an
additional equipotential bonding line�
Unless a low-current connection between the pyrometer and the electrical cabinet
of the installation can be ensured, the pyrometer must be mounted in an insulated
manner to avoid equalising currents� The shielding must be connected to the
plant’s earthing system�
If the pyrometer is installed without an insulator and with out potential
equalisation, the interference voltage may not exceed 32V�