E-4
Example Programs
Models 2510 and 2510-AT User’s Manual
Setpoint tolerance temperature sweep
Overview
As discussed in detail in
, the setpoint tolerance feature allows the unit to gener-
ate an SRQ when the measured temperature is within a certain percent tolerance of the set-
point for a specific number of readings. This feature can be used to generate temperature
sweeps, as we will now discuss.
demonstrates the basic method for generating temperature sweeps. Once the
temperature stabilizes at the Initial Setpoint, a step change is made in the setpoint setting.
Once the temperature is within the programmed setpoint tolerance, the Model 2510 gener-
ates an SRQ to the computer to indicate that the new setpoint has been reached. The pro-
cess is then repeated for as many steps in the temperature sweep as required.
Figure E-1
Setpoint tolerance temperature sweep
SRQ
SRQ
SRQ
SRQ
Change Setpoint
Change Setpoint
Change Setpoint
Step 1
Step 2
Initial
Setpoint
Final
Setpoint