Connect the instrument 

Important test system safety information 

This product is sold as a stand-alone instrument that may become 
part of a system that could contain hazardous voltages and energy 

sources. It is the responsibility of the test system designer, integrator, 

installer, maintenance personnel, and service personnel to make sure 
the system is safe during use and is operating properly. 

 You must also realize that in many test systems a single fault, such as 

a software error, may output hazardous signal levels even when the 
system indicates that there is no hazard present. 

It is important that you consider the following factors in your system 
design and use: 

 

The international safety standard UL 61010-1 defines voltages as 

hazardous if they exceed 30 V RMS and 42.4 V peak, or 60 V DC 

for equipment rated for dry locations. Keithley Instruments 
products are only rated for dry locations. 

 

Read and comply with the specifications of all instruments in the 

system. The overall allowed signal levels may be constrained by 
the lowest rated instrument in the system. For example, if you are 

using a 500 V DC power supply with a 300 V DC rated switch, 

the maximum allowed voltage in the system is 300 V DC. 

 

Make sure any test fixture connected to the system protects the 

operator from contact with hazardous voltages, hot surfaces, and 
sharp objects. Use shields, barriers, insulation, safety interlocks, 

and so on to accomplish this. 

 

Cover the device under test (DUT) to protect the operator from 

flying debris in the event of a system or DUT failure. 

 

Double-insulate all electrical connections that an operator can 
touch. Double insulation ensures that the operator is still 

protected even if one insulation layer fails. Refer to IEC 61010-1 

for specific requirements. 

 

Make sure all connections are behind a locked cabinet door or 
other barrier. This protects the system operator from accidentally 

removing a connection by hand and exposing hazardous voltages. 

Содержание System SourceMeter 2602B

Страница 1: ...Models 2601B 2602B and 2604B System SourceMeter Instruments Quick Start Guide...

Страница 2: ...ent Category I and Measurement Category II as described in the International Electrotechnical Commission IEC Standard IEC 60664 Most measurement control and data I O signals are Measurement Category I...

Страница 3: ...be used as shield connections for measuring circuits NOT as protective earth safety ground connections If you are using a test fixture keep the lid closed while power is applied to the device under t...

Страница 4: ...r installation into a computer should never require cleaning if handled according to instructions If the board becomes contaminated and operation is affected the board should be returned to the factor...

Страница 5: ...nce troubleshooting and programming commands The documentation is in PDF format on the CD ROM that is included with the instrument If you do not have Adobe Reader to view the files you can download a...

Страница 6: ...Unpack and inspect the instrument To unpack and inspect the instrument 1 Inspect the box for damage 2 Open the top of the box...

Страница 7: ...documentation accessories and CD ROMs 4 Carefully lift the instrument out of the box 5 Remove the packaging insert 6 Inspect the instrument for any obvious signs of physical damage Report any damage...

Страница 8: ...or the Models 2601B and 2602B 4 Power line cord 5 One 8 pin screw terminal connector for the Model 2601B Two 8 Pin screw terminal connectors for the Models 2602B and 2604B 6 Series 2600B Product Infor...

Страница 9: ...ed for dry locations Read and comply with the specifications of all instruments in the system The overall allowed signal levels may be constrained by the lowest rated instrument in the system For exam...

Страница 10: ...2600B can be used on a bench or in a rack Please see the instructions that came with your rack mount kit if you are installing the Series 2600B in a rack Note that the air intakes for the fan are loc...

Страница 11: ...ne power 1 Make sure the front panel power switch is in the off 0 position 2 Connect the socket of the supplied power cord to the power connection on the rear panel 3 Connect the plug of the power cor...

Страница 12: ...odel 2602B the display will show KEITHLEY Model 2602B Line frequency detection and other startup checks The entire power up process takes approximately 30 seconds to complete When initialization is co...

Страница 13: ...est DUT for this test Step 1 Set source function range and level If the instrument has two channels Models 2602B or 2604B and is in dual channel display mode press the DISPLAY button once to show only...

Страница 14: ...e navigation wheel to set the source value to 20 0000 V and then press the navigation wheel to enter the selection and exit EDIT mode Step 2 Set the source limit 1 Press the LIMIT key You will see a b...

Страница 15: ...the Series 2600B automatically selects the best range for the measured value You will briefly see the display shown below and then the main display screen reappears Step 4 Turn the output on Turn the...

Страница 16: ...output indicator LED turns off These steps confirm basic functionality of your Model 2601B Model 2602B or Model 2604B instrument Turn the instrument power OFF now For a basic understanding of the of t...

Страница 17: ...ept Model 2400 SCPI commands For more information about loading and running the 2400 personality script see Model 2400 emulation in the Series 2600B Reference Manual What do I do if I lost or formatte...

Страница 18: ......

Отзывы: