Appendix C: LAN concepts and settings
Series 2600B System SourceMeter® Instrument Reference Manual
C-8
2600BS-901-01 Rev. B / May 2013
5. Select
Use the following IP address
. The option for "Use the following DNS server addresses"
is automatically selected.
6. Set the IP address. If the IP address and subnet mask fields:
•
Contain values:
Record the IP address, subnet mask, default gateway, and DNS servers to use in
Step
3: Configure the instrument's LAN settings
(on page C-8).
•
Are blank:
In the IP address field, enter 192.168.1.100. In the subnet mask field, enter 255.255.255.0.
These will be used to configure the LAN settings of the instrument.
1. Click
OK
to close the Internet Protocol Version 4 (TCP/IPv4) Properties dialog box.
2. Click
OK
to close the Local Area Connection Properties dialog box.
3. Close the Network Connections window.
Step 3: Configure the instrument's LAN settings
These steps assume that you are making all the settings in the order shown here. If you only change
one or a few settings, be aware that you need to apply the settings before they will be in effect. To
apply the settings, from the
LAN CONFIG
menu, select
APPLY_SETTINGS > YES
, and then press
the
ENTER
key.
To configure the Series 2600B using the front panel:
1. Press the
MENU
key to display the MAIN MENU.
2. Use the navigation wheel
to select
LAN
. The LAN CONFIG menu is displayed.
3. Change the IP address assignment method:
a. Select
CONFIG > METHOD > MANUAL
, and then press the
ENTER
key.
b. Press the
EXIT (LOCAL)
key once to return to the LAN CONFIG menu.
4. Enter the IP address using the LAN CONFIG menu:
a. Select
CONFIG > IP-ADDRESS
.
b. Refer to the recorded computer's IP address (
Step 1: Identify and record the existing IP configuration
(on page C-2)). A portion of the computer's IP address is used as a base for the instrument's unique ID.
Only the last three numbers (after the last decimal point) of the IP address will differ between the
computer and the instrument. If the subnet mask is 255.255.255.0, the last three digits can be any value
from 1 to 255.
For example, the Internet Protocol (TCP/IP) Properties dialog box shows that the computer's IP address
is 192.168.1.100 (see the figure titled "Internet protocol (TCP/IP) Properties dialog box" in
Step 2:
Disable DHCP to use the computer's existing IP address
(on page C-4)). A unique IP address for the
instrument might be 192.168.001.101.
The instrument’s IP address can have leading zeros, but the computer’s IP address cannot.
c.
Use the navigation wheel
to select and enter an appropriate IP address for the instrument. Be sure
to record the instrument’s IP address to use in
Step 5: Access the instrument's web page
(on page C-
d. Press
ENTER
key or navigation wheel
to confirm the changes.
e. Press the
EXIT (LOCAL)
key twice to return to the LAN CONFIG menu.