Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-358
2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].mode
This attribute defines the trigger operation and detection mode. This attribute is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
TSP-Link trigger N reset
Not saved
0 (tsplink.TRIG_BYPASS)
Usage
mode
= tsplink.trigger[
N
].mode
tsplink.trigger[
N
].mode =
mode
mode
The trigger mode
N
The trigger line (1 to 3)
Details
This attribute controls the mode in which the trigger event detector and the output trigger generator operate on
the given trigger line.
The setting for
mode
can be one of the following values:
Mode
Number
value
Description
tsplink.TRIG_BYPASS
0
Allows direct control of the line as a digital I/O line.
tsplink.TRIG_FALLING
1
Detects falling-edge triggers as input. Asserts a TTL-low pulse
for output.
tsplink.TRIG_RISING
2
If the programmed state of the line is high, the
tsplink.TRIG_RISING
mode behaves similarly to
tsplink.TRIG_RISINGA
.
If the programmed state of the line is low, the
tsplink.TRIG_RISING
mode behaves similarly to
tsplink.TRIG_RISINGM
.
Use
tsplink.TRIG_RISINGA
if the line is in the high output
state.
Use
tsplink.TRIG_RISINGM
if the line is in the low output
state.
tsplink.TRIG_EITHER
3
Detects rising- or falling-edge triggers as input. Asserts a
TTL-low pulse for output.
tsplink.TRIG_SYNCHRONOUSA
4
Detects the falling-edge input triggers and automatically
latches and drives the trigger line low.
tsplink.TRIG_SYNCHRONOUS
5
Detects the falling-edge input triggers and automatically
latches and drives the trigger line low. Asserts a TTL-low pulse
as an output trigger.
tsplink.TRIG_SYNCHRONOUSM
6
Detects rising-edge triggers as an input. Asserts a TTL-low
pulse for output.
tsplink.TRIG_RISINGA
7
Detects rising-edge triggers as input. Asserts a TTL-low pulse
for output.
tsplink.TRIG_RISINGM
8
Edge detection as an input is not available. Generates a
TTL-high pulse as an output trigger.
Содержание System SourceMeter 2601B