Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-226
2600BS-901-01 Rev. B / May 2013
Also see
Output-off states
(on page 2-77)
smuX.source.offlimitY
(on page 7-226)
smuX.source.offmode
(on page 7-226)
smuX.source.output
(on page 7-228)
smuX.source.offlimitY
This attribute sets the limit (current or voltage) used when the source-measure unit (SMU) is in normal output-off
mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
Current: 1e-3 (1 mA)
Voltage: 40 (40 V)
Usage
sourceLimit
= smu
X
.source.offlimit
Y
smu
X
.source.offlimit
Y
=
sourceLimit
sourceLimit
Set to the limit to be used when the SMU is in normal output-off mode
X
Source-measure unit (SMU) channel (for example,
smua.source.offlimiti
applies to SMU channel A)
Y
SMU source function (
v
= voltage,
i
= current)
Details
Setting the current limit to lower than 1 mA may interfere with operation of the contact check function. See
smu
X
.contact.check()
and
smu
X
.contact.r()
for details.
Example
smua.source.offlimiti = 10e-3
Changes the normal output-off
mode limit to 10 mA for SMU
channel A.
Also see
smuX.contact.check()
(on page 7-193)
smuX.contact.r()
(on page 7-194)
smuX.source.offfunc
smuX.source.offmode
(on page 7-226)
smuX.source.offmode
This attribute sets the source output-off mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Saved setup
0 (smuX.OUTPUT_NORMAL)