3-14 Test Script Processor Interaction
Series 2600 System SourceMeters User’s Manual
2600S-900-01 Rev. A / May 2006
Return to
How do I use other programs?
Reference
Refer to the LabVIEW and Visual Basic documentation for details
on using those programs.
Using LabVIEW
Source step example
The source step example is shown in
. The example steps the source
voltage from 1V to 10V in 1V steps, measures the current at each step, and stores
the result in a table. Once all steps are completed, the stored table current values
are printed.
The Series 2600 command sequence:
localnode.prompts = 0
-- Disable prompts.
data_table = {}
-- Clear table.
reset()
-- Reset Series 2600.
smua.source.output = 1
-- Turn on source output.
for i=1,10 do
-- Loop for 10 steps.
smua.source.levelv = i
-- Set source level.
mi = smua.measure.i()
-- Take a current measurement.
table.insert(data_table,mi)
-- Store measurement in table.
end
-- End of loop.
smua.source.output = 0
-- Turn off output.
print(table.concat(data_table,’,’))
-- Print table of current values.
reset()