Section 12: Instrument Control Library
Series 2600 System SourceMeter® Instruments Reference Manual
Remarks
• This function generates and activates new calibration constants for the given range. The
positive and negative polarities of the instrument must be calibrated separately. Use a
positive value fo
r
range
to
calibrate the positive polarity and a negative value fo
r
range
to
calibrate the negative polarity.
• Typically the two calibration points used will be near zero for calibration point 1 and 90%
of full scale for calibration point 2.
• All four measurements (
cp1measured
,
cp1reference
,
cp2measured
, and
cp2reference
)
must
be made with the active calibration set. Corruption of the
calibration constants may result if this is not heeded.
•
The new calibration constants will be activated immediately but they will not be
written to non-volatile storage. Use
smuX.cal.save
to commit the new constants to
nonvolatile storage. The active calibration constants will stay in effect until the
instrument is power cycled or a calibration set is loaded from non-volatile storage
with the
smuX.cal.restore
function.
•
This function will be disabled until a successful call to
smuX.cal.unlock
is made.
Details
See
"Calibration".
Also see
,
,
smuX.measure.count
X = SMU channel (a or b)
Attribute
Number of measurements performed when a measurement is requested.
Usage
count = smuX.measure.count
-- Reads measure count.
smuX.measure.count = count
-- Writes measure count.
count
Number of measurements.
Remarks
• This attribute controls the number of measurements taken any time a measurement is
requested. When using a reading buffer with a measure command, the
count
also
controls the number of readings to be stored.
•
The
reset
function sets the measure count to 1.
Details
See “
Also see
,
Example
Sets measure count for SMU A:
smua.measure.count = 10
smuX.measure.delay
X= SMU channel (a or b)
(Models 2635 and 2636 only)
Attribute
Controls the measurement delay.
Usage
mdelay = smuX.measure.delay
-- Read the measure delay.
smuX.measure.delay = mdelay
-- Write the measure delay.
Set
mdelay
to one of the following values:
0
or
smuX.DELAY_OFF
No delay.
-1
or
smux.DELAY_AUTO
Automatic delay value.
user_value
Set user delay value.
Remarks
• This attribute allows for additional settling time before taking a measurement.
• The default is
smuX.DELAY_AUTO.
• The
smuX.DELAY_AUTO
setting causes a current range-dependent delay to be inserted
when a current measurement function is called.
• If
smuX.measure.count
is greater than 1, the measurement delay is only inserted
before the first measurement.
•
mdelay
can be set to a specific user-defined value that sets the delay that is used
regardless of range.
• Use
smuX.source.delay
when measuring your source (source read back) and
smuX.measure.delay
when stepping voltage and measuring current.
Also see
Example
smua.measure.delay = 0.010
Sets the measurement delay to 10mS.