Models 2611/2612
System SourceMeter
®
Specifications
10
ȝ
A range:
<500
ȝ
s typical.
1
ȝ
A range:
<2ms typical.
100nA range:
<20ms typical.
DC floating voltage:
Output can be floated up to ±250VDC from chassis ground.
Remote sense operating range:
1
Maximum voltage between HI and SENSE HI = 3V. Maximum voltage between LO and
SENSE LO = 3V.
Voltage output headroom:
200V range:
Max. output voltage = 202.3V; total voltage drop across source leads (maximum 1
ȍ
per source lead).
20V range:
Max. output voltage = 23.3V; total voltage drop across source leads (maximum 1
ȍ
per source lead).
Over temperature protection:
Internally-sensed temperature overload puts unit in standby mode.
Voltage source range change overshoot:
Overshoot into a 200k load, 20MHz BW, 300mV typical.
Current source range change overshoot:
<5% of larger range + 300mV/Rload + 60nA typical (see “Current source output
settling time” for additional test conditions).
Pulse Specifications
Region
Maximum current limit
Maximum pulse width
7
Maximum duty cycle
8
1
100mA at 200V
DC, no limit
100%
1
1.5A at 20V
DC, no limit
100%
2
1A at 180V
8.5ms
1%
3
9
1A at 200V
2.2ms
1%
4
10A at 5V
1ms
2.2%
Minimum programmable pulse width:
7
200
ȝ
s. NOTE: Minimum pulse width for settled source at a given I/V
output and load can be longer than 200
ȝ
s. See note 10 for typical settling times.
10
7 Times measured from the start of pulse to the start of off-time:
Pulse Level
Bias Level
Start t
on
Start t
off
90%
10%
t
on
t
off
10%
8 Thermally limited in sink mode (quadrants 2 and 4) and ambient temperatures above 30°C. See power equations in the
Reference Manual for more information.
9 Voltage source operation with 1.5A current limit.
10 Typical performance for minimum settled pulse widths:
Source value
Load
Source settling (% of range)
Min. pulse width
5V 0.5
ȍ
1% 300µs
20V 200
ȍ
0.2%
200µs
180V 180
ȍ
0.2%
5ms
200V (1.5A limit)
200
ȍ
0.2%
1.5ms
100 mA
200
ȍ
1%
200µs
1A 20
ȍ
1%
500µs
1A 180
ȍ
0.2%
5ms
10A 0.5
ȍ
0.5%
300µs
Typical tests were performed using remote operation, 4W sense, Keithley 2600 ban cables and best fixed
measurement range. For more information on pulse scripts, see the Series 2600 Reference Manual.