16-18 Calibration
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
7. Repeat steps a through v for the 3A and 10A ranges (Model 2601/2602) or
1.5A and 10A ranges (Model 2611/2612). Compute the current reading
from the DMM voltage reading and characterized 0.5
Ω
resistance value: I =
V/R.
8. Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO
Figure 16-3
Connections for current calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
!
!
!
UL
C
US
LISTED
SourceMeter
4ZA4
CAT I
S
S
LO
LO
HI
HI
G G G
G
CHANNEL B
!
R
Digital Multimeter
Input HI
Input LO
Model 2602/2612 Channel A
Connections Shown
LO
HI
0.5
W
250W
Resistor
Sense
Terminals
Load
Terminals
(2601/2602 3A and 10A ranges; 2611/2612 1.5A and 10A ranges)