12-88 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
smuX.source.levelY
X = SMU channel (a or b)
Y = SMU measure function (v or i) where v = voltage, i = current
Attribute
Source levels.
Usage
sourceval = smuX.source.levelY
-- Reads source value.
smuX.source.levelY = sourceval
-- Writes source value.
sourceval
Set 2601/2602 voltage from 0 to ±40 (volts).
Set 2601/2602 current from 0 to ±3 (amps).
Set 2611/2612 voltage from 0 to ±200 (volts).
Set 2611/2612 current from 0 to ±1.5 (amps).
Remarks
• This attribute configures the source level of the voltage or current source.
• If the source is configured as a voltage source and the output is on, the new
smuX.source.levelv
setting will be sourced immediately. If the output is off or if
the source is configured as a current source, the voltage level will be sourced when
the source is configured as a voltage source and the output is turned on.
• If the source is configured as a current source and the output is on, the new
smuX.source.leveli
setting will be sourced immediately. If the output is off or if
the source is configured as a voltage source, the current level will be sourced when
the source is configured as a current source and the output is turned on.
• The sign of
sourceval
dictates the polarity of the source. Positive values generate
positive voltage or current from the high terminal of the source relative to the low
terminal. Negative values generate negative voltage or current from the high
terminal of the source relative to the low terminal.
• The
reset
function sets the source levels to 0V and 0A.
Details
(basic operation)
Also see
Example
Sets V-source to 1V for SMU A:
smua.source.levelv = 1
smuX.source.limitY
X = SMU channel (a or b)
Y = SMU measure function (v or i) where v = voltage, i = current
Attribute
Compliance limits.
Usage
limit = smuX.source.limitY
-- Reads compliance limit.
smuX.source.limitY = limit
-- Writes compliance limit.
limit
2601/2602 voltage compliance from 0 to ±40 (volts).
2601/2602 current compliance from 0 to ±3 (amps).
2611/2612 voltage compliance from 0 to ±200 (volts).
2611/2612 current compliance from 0 to ±1.5 (amps).
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......