12-62 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
smuX.cal.password
X = SMU channel (a or b)
Attribute
Password to enable calibration.
Usage
smuX.cal.password = newpassword
newpassword
The new password (string).
Remarks
• A new password can only be assigned when calibration has been unlocked.
• The calibration password is write-only and cannot be read.
Details
(calibration).
Also see
Example
Assign a new calibration password for SMU A:
smua.cal.password = "LetMeIn"
smuX.cal.polarity
X = SMU channel (a or b)
Attribute
Control which calibration constants are used for all subsequent measurements.
Usage
calpolarity = smuX.cal.polarity
-- Reads cal polarity.
smuX.cal.polarity = calpolarity
-- Writes cal polarity.
Set
calpolarity
to one of the following values:
0
or
smuX.CAL_AUTO
Automatic
polarity
detection.
1
or
smuX.CAL_POSITIVE
Measure with positive polarity calibration constants.
2
or
smuX.CAL_NEGATIVE
Measure with negative polarity calibration constants.
Remarks
• This attribute controls which polarity calibration constants are used to make all
subsequent measurements. This attribute does not affect the
smuX.measure.calibrateY
or the
smuX.source.calibrateY
function. The
polarity for those commands are dictated by the range parameter given to the
command.
• The measurement calibration commands require the measurements provided to
have been made using the polarity being calibrated. When the calibration points are
sufficiently far away from zero the desired polarity constants are inherently used
when making those measurements. When measuring near zero, it is possible for the
measurement to be made using the calibration constants from either polarity without
knowing which was used. Setting this attribute to positive or negative forces
measurements to be made using the calibration constants for a given polarity rather
than basing the choice on the raw measurement data.
• This attribute can only be set to positive or negative when calibration is unlocked.
This attribute will automatically be set back to CAL_AUTO when calibration is
locked.
Details
(calibration).
Also see
Содержание SourceMeter 2600 Series
Страница 507: ...AppendixA Specifications Appendix A topics Model 2600S Specifications page A 2 2600S 901 01 Rev A May 2006...
Страница 575: ...AppendixF DisplayCharacterCodes Appendix F topics Introduction page F 2...
Страница 592: ......
Страница 593: ......