10-10 Digital I/O Port
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Remote control of Output Enable
Use one of these commands to control Output Enable action:
smuX.source.outputenableaction = smuX.OE_NONE
smuX.source.outputenableaction = smuX.OE_OUTPUT_OFF
When set to
smuX.OE_NONE
, the Model 2601/2602 SourceMeter will take no
action when the Output Enable line goes low. When set to
smuX.OE_OUTPUT_OFF
, the SourceMeter will turn its output off as if the
smuX.source.output = smuX.OUTPUT_OFF
command had been received.
The SourceMeter will not automatically turn its output on when the Output Enable
line returns to the high state. For example, the following command activates the
Output Enable for Channel A:
smua.source.outputenableaction = smua.OE_OUTPUT_OFF
Figure 10-3
Using Model 2601/2602 Output Enable
OE
(pin 24)
+5V
(pin 23)
Switch (Lid Closed)
Digital I/O
Model 2601/2602
SourceMeter
Test Fixture
A. OUTPUT can be turned on.
INT
(pin 24)
+5V
(pin 23)
Switch (Lid Open)
Model 2601/2602
SourceMeter
Test Fixture
B. OUTPUT cannot be turned on.
Digital I/O