B-8
Command Reference
active source function/range point will be calibrated using the corresponding
parameter.
Note that parameters for a given source function and range may be sent in
any order; however, once one :CAL:PROT:SOUR command executes, the
other three must also be executed, or an error number -200, “Execution
error” will occur when the :CAL:PROT:SAVE command is processed at the
end of the calibration procedure.
Because the source is calibrated for both positive and negative values, two
zero calibration points are required. The Model 2425 automatically uses the
appropriate zero parameter based on whether the source polarity is positive
or negative.
Example
:CAL:PROT:SOUR 2
Calibrate 2V source range.
Table B-3
Recommended :CALibration:PROTected:SOURce parameter ranges
Source
range
First parameter
(negative full scale)
Second parameter
(negative zero)
Third parameter
(positive full scale)
Fourth parame-
ter (positive zero)
0.2V
2V
20V
100V
10
μ
A
100
μ
A
1mA
10mA
100mA
1A
3A
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-90 to -110
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -11E-3
-90E-3 to -110E-3
-0.9 to -1.1
-2.7 to -3.15
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-1 to +1
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to +3E-2
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+90 to +110
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+2.7 to +3.15
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-1 to +1
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to +3E-2
Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75% of full scale to 150% of full
scale for full-range parameters, but ranges listed above should be observed for optimum calibration.
Содержание SourceMeter 2425
Страница 1: ...Model 2425100W SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Страница 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Страница 15: ......
Страница 16: ...1 PerformanceVerification...
Страница 35: ...1 20 Performance Verification Model 2425 Service Manual...
Страница 36: ...2 Calibration...
Страница 62: ...3 RoutineMaintenance...
Страница 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Страница 66: ...4 Troubleshooting...
Страница 84: ...5 Disassembly...
Страница 92: ...6 ReplacementParts...
Страница 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Страница 114: ...A Specifications...
Страница 116: ...B CommandReference...
Страница 128: ...C CalibrationPrograms...
Страница 139: ......
Страница 141: ......