background image

   

10   

INST:SEL   CH3   
SOURCE:VOLT   5V   
SOURCE:CURR   1.5A   
   
INST:SEL   CH1   
SOURCE:OUTP:ENAB   ON   
SOURCE:OUTP   ON   
..<insert   desired   amount   of   delay   between   channel   1   turn   on   and   channel   2   turn   on   using   
the   appropriate   command   syntax   of   the   programming   environment>…   
INST:SEL   CH2   
SOURCE:OUTP:ENAB   ON   
..<insert   desired   amount   of   delay   between   channel   2   turn   on   and   channel   3   turn   on   using   
the   appropriate   command   syntax   of   the   programming   environment>…   
INST:SEL   CH3   
SOURCE:OUTP:ENAB   ON   
…   
<perform   tests   on   the   load>   
…   
INST:SEL   CH3   
SOURCE:OUTP:ENAB   OFF   
..<insert   desired   amount   of   delay   between   channel   3   turn   off   and   channel   2   turn   off   
using   the   appropriate   command   syntax   of   the   programming   environment>…   
INST:SEL   CH2   
SOURCE:OUTP:ENAB   OFF   
..<insert   desired   amount   of   delay   between   channel   2   turn   off   and   channel   1   turn   off   
using   the   appropriate   command   syntax   of   the   programming   environment>…   
INST:SEL   CH1   
SOURCE:OUTP:ENAB   OFF   

Содержание Series 2200

Страница 1: ...o select Enable Channels 3 Press Enter to enter the submenu 4 Use the arrow keys or the navigation wheel to select Enable Disable CH2 5 Press Enter to enter the submenu 6 Use the arrow keys or the nav...

Страница 2: ...channel 1 and 2 Front Panel Operation STEP 1 Configure the Instrument to Series Combination 1 Press Menu 2 Use the arrow keys or the navigation wheel to select Combine CH1 CH2 3 Press Enter to enter t...

Страница 3: ...t S e n s e CHANNEL 2 O u t S e n e O u t S e n s e CHANNEL 3 O u t S e n s e REAR PANEL SOURCE HIGH SOURCE LOW SENSE HIGH SENSE LOW Figure 1 Connecting a Device in Series Combination on the Front Pa...

Страница 4: ...limit 12 Press Enter to set the current of the combined channel STEP 3 Turning on the Output 13 Press Output On Off Remote Operation The following SCPI command example configures the power supply for...

Страница 5: ...2 Press V Set 3 Use the arrow keys numeric keypad or the navigation wheel to enter a voltage output e g 10V 4 Press Enter to set the voltage of the combined channel 5 Press I Set 6 Use the arrow keys...

Страница 6: ...will change the other channel by the desired ratio Remote Operation The following SCPI command example sets CH1 voltage output to 10V the current limit to 1 5A CH2 voltage output to track half of CH1...

Страница 7: ...n the Front Panel and the Rear Panel Turning On Multiple Channels Sequentially The Series 2200 multi channel power supplies have all channels enabled in its default state The following example demonst...

Страница 8: ...nable Disable CH1 26 Press Enter to enter the submenu 27 Use the arrow keys or the navigation wheel to select Enable CH1 default 28 Press Enter to enable channel 1 29 Press Esc a couple of times to re...

Страница 9: ...e arrow keys or the navigation wheel to select Disable CH2 61 Press Enter to disable channel 2 and turn off channel 2 s output voltage 62 Press Esc a couple of times to return to the main display STEP...

Страница 10: ...nnel 3 turn on using the appropriate command syntax of the programming environment INST SEL CH3 SOURCE OUTP ENAB ON perform tests on the load INST SEL CH3 SOURCE OUTP ENAB OFF insert desired amount of...

Отзывы: