Section 2: S540 site preparation and installation
S540 Power Semiconductor Test System Administrative Guide
2-24
S540-924-01 Rev. D / January 2019
Cable harness routing
The S540 has a dual-door rear portal for output leads. Cables can be routed from the cable harness
at the rear portal to either the right or left of the cabinet.
Figure 27: S540 cabinet rear doors and cable portal
You can mount cable support arms on either side of the cabinet to suspend cables off the floor and
prevent a tripping hazard.
Floor plan
The following floor plan information is for the system cabinet only. Be sure to place the cabinet a
minimum distance of 15 cm (6 in.) up to a maximum distance of 122 cm (48 in.) from the prober.
Refer to the documentation for the prober or other test-fixture equipment to determine its floor space
requirements.