S46L Microwave Switch System Relay Installation Guide

Relays A through D installation

1.  Remove the four (4) screws securing the cover plate over the mounting hole where the new relay 

will be installed (see 

Figure 4

).

Figure 4: Remove mounting hole cover plate

2.

Assemble the supplied ribbon cable to the relay (use 

Figure 5

 as a guide):

    

 

 

Figure 5: Connecting control wires to relays A through D 

3.

Install the relay in the mounting hole, and then secure it with the four (4) screws removed in step 1.

4.

Plug the ribbon connector into the appropriate circuit board connector (see 

Figure 3

).

Cover plate