Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-121
:TRACe:STATistics:MAXimum?
This command returns the maximum reading value in the reading buffer.
Type
Affected by
Where saved
Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:STATistics:MAXimum?
:TRACe:STATistics:MAXimum? <bufferName>
<bufferName>
A string that indicates the reading buffer; the default buffers
(
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no
buffer is specified, this parameter defaults to
defbuffer1
Example
TRACe:STAT:MAXimum?
Returns the maximum reading value in the default
buffer,
defbuffer1
.
TRACe:STAT:MAXimum? "testData"
Returns the maximum reading value in the user-defined
buffer
testData
.
Also see
(on page 3-10)
(on page 3-27)
(on page 6-114)
(on page 6-119)
(on page 6-120)
(on page 6-121)
(on page 6-122)
(on page 6-123)
:TRACe:STATistics:MINimum?
This command the minimum reading value in the reading buffer.
Type
Affected by
Where saved
Default value
Query only
Recall settings
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
:TRACe:STATistics:MINimum?
:TRACe:STATistics:MINimum? <bufferName>
<bufferName>
A string that indicates the reading buffer; the default buffers
(
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no
buffer is specified, this parameter defaults to
defbuffer1