Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-242
DMM7510-901-01 Rev. B / May 2015
status.operation.enable
This attribute sets or reads the contents of the Operation Event Enable Register of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
status.preset()
Nonvolatile memory
0
Usage
operationRegister
= status.operation.enable
status.operation.enable = o
perationRegister
operationRegister
The status of the operation status register
Details
This command sets or reads the contents of the Enable register of the Operation Event Register.
When one of these bits is set, when the corresponding bit in the Operation Event Register or
Operation Condition Register is set, the OSB bit in the Status Byte Register is set.
Example
-- decimal 20480 = binary 0101 0000 0000 0000
status.operation.enable = operationRegister
Sets the 12 and 14 bits of the operation
status enable register using a decimal
value.
Also see