Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-126
DMM7510-901-01 Rev. B / May 2015
Also see
(on page 3-64)
dmm.digitize.analogtrigger.pulse.polarity
(on page 8-78)
dmm.measure.analogtrigger.mode
(on page 8-121)
dmm.measure.analogtrigger.pulse.condition
(on page 8-122)
dmm.measure.analogtrigger.pulse.polarity
(on page 8-125)
dmm.measure.analogtrigger.pulse.width
(on page 8-126)
dmm.measure.analogtrigger.pulse.width
This attribute defines the threshold value for the pulse width.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
1 ms
Usage
value
= dmm.measure.analogtrigger.pulse.width
dmm.measure.analogtrigger.pulse.width =
value
value
The threshold value for the pulse width: 1 µs to 40 ms
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This option is only available when the analog trigger mode is set to pulse.
This option sets either the minimum or maximum pulse width that generates an analog trigger event.
The value of pulse condition determines whether this value is interpreted as the minimum or