Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-219
:TRIGger:LOAD "ConfigList"
This command loads a predefined trigger model configuration that uses source and measure configuration lists.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>
:TRIGger:LOAD "ConfigList", "<measureConfigList>", "<sourceConfigList>", <delay>,
"<bufferName>"
:TRIGger:LOAD "ConfigList", "<measureConfigList>", <delay>, "<bufferName>",
<readingBlock>
<measureConfigList>
A string that contains the name of the measurement configuration list to
use
<delay>
The delay time before each measurement (167 ns to 10 ks); default is 0 for
no delay
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is
specified,
defbuffer1
is used
<readingBlock>
Define a measure or digitize block for the trigger model; options are:
•
ACTive
: Add a measure or digitize block to the trigger model based
on the active function; if no option defined,
ACTive
is used
•
MEASure
: Adds a measure block to the trigger model
•
DIGitize
: Adds a digitize block to the trigger model
Details
This trigger model template incorporates a configuration list. You must set up the configuration lists
before loading the trigger model.
You can also set a delay and change the reading buffer.
The out line of the EXT TRIG IN/OUT rear-panel terminal is asserted at the end of each
measurement.
After selecting a trigger model template, you can view the trigger model blocks in a graphical format
by pressing the front-panel
MENU
key and under Trigger, selecting
Configure
. You can also add or
delete blocks and change trigger model settings from this screen. You can use the
TRIGger:BLOCk:LIST?
command to view the trigger model blocks in a list format.