background image

 

4-8

Cabling and Wiring

 

Note:  

 

Each EXP-16, EXP-16/A, or EXP-GP expansion board is 

associated with an analog input channel on the DAS-800 Series board. 
You specify the associated on-board channel by setting a jumper on the 
expansion board. Make sure that you use a unique jumper setting for each 
expansion board you are using. Refer to your expansion board 

 

documentation for more information. 

You can use the J2 connector on the last expansion board in the chain to 
attach an STA-08 / STA-08PGA screw terminal accessory. This allows 
you to access the remaining functions of the DAS-800 Series board. 

Figure 4-7 illustrates how to connect two EXP-16 expansion boards, one 
EXP-GP expansion board, and an STA-08 screw terminal accessory to a 
DAS-800 board. 

 

Figure 4-7.  Connecting Multiple Expansion Boards

 

Note:  

 

You can also attach up to four MB-02 backplanes to a DAS-800 

Series board using an STA-SCM8 screw terminal accessory and up to 
four C-2600 cables. Refer to the 

 

MB Series User’s Guide

 

 for 

 

more information. 

DAS-800 

Board

EXP-16

EXP-16

EXP-GP

STA-08

J1

J2

J1

J2

J1

J2

J1

J2

Содержание DAS-800 series

Страница 1: ...DAS 800 Series User s Guide A G R E A T E R M E A S U R E O F C O N F I D E N C E...

Страница 2: ...trademarks of their respective companies Information furnished by Keithley Instruments is believed to be accurate and reliable However Keithley Instruments assumes no responsibility for the use of su...

Страница 3: ...DAS 800 Series User s Guide Revision D June 2002 Part Number 86870...

Страница 4: ...otherwise in accordance with the instructions therefore Keithley does not warrant that operation of the Keithley Software will be uninterrupted or error free and or that the Keithley Software will be...

Страница 5: ...nt Pieters Leeuw 02 363 00 40 Fax 02 363 00 64 CHINA Yuan Chen Xin Building Room 705 12Yumin Road Dewai Madian Beijing 100029 8610 6202 2886 Fax 8610 6202 2892 FINLAND Tiet j ntie 2 02130 Espoo Phone...

Страница 6: ...nstallation Category I and Installation Category II as described in the International Electrotechnical Commission IEC Standard IEC 60664 Most mea surement control and data I O signals are Installation...

Страница 7: ...ld refer to the operating instructions located in the manual The symbol on an instrument shows that it can source or measure 1000 volts or more including the combined effect of normal and common mode...

Страница 8: ...s 2 14 Data Transfer 2 16 Digital I O Features 2 17 8254 Counter Timer Circuitry 2 17 Interrupts 2 25 Power 2 26 3 Setup and Installation Installing and Configuring DriverLINX for DAS 800 Series Board...

Страница 9: ...nal to a Differential Input 4 10 Connecting Digital Signals 4 12 Connecting Counter Timer I O Signals 4 12 Synchronizing Conversions on Multiple Boards 4 13 5 DriverLINX Analog I O Panel Test Panel Ap...

Страница 10: ...re 4 2 Main I O Connector on a DAS 801 or DAS 802 Board 4 3 Figure 4 3 Attaching an STC 37 Screw Terminal Connector 4 4 Figure 4 4 Attaching an STA 08 STA 08PGA Screw Terminal Accessory 4 5 Figure 4 5...

Страница 11: ...er Limitations 2 27 Table 3 1 I O Address Map 000H to 3FFH 3 11 Table 3 2 Interrupt Levels 3 15 Table 7 1 Troubleshooting Information 7 4 Table A 1 DAS 800 DAS 801 and DAS 802 Specifications A 9 Table...

Страница 12: ...s of DAS 800 Series boards including a description of supported software and accessories Chapter 2 provides a more detailed description of the analog input digital I O and counter timer I O features o...

Страница 13: ...keep the following in mind References to DAS 800 Series boards apply to the DAS 800 DAS 801 and DAS 802 boards When a feature applies to a particular board that board s name is used References to Micr...

Страница 14: ...mmended The DAS 800 Series includes the DAS 800 DAS 801 and DAS 802 boards The major features of DAS 800 Series boards are as follows Eight analog input channels single ended on the DAS 800 single end...

Страница 15: ...Using DriverLINX With Your Hardware Configuring the DAS 800 Series manuals Supporting Software The following software is available for operating DAS 800 Series boards DAS 800 Series standard software...

Страница 16: ...reference and information specific to the DAS 800 Series hardware DAS 800 Series utilities The following utilities are provided as part of both the DAS 800 Series standard software package DriverLINX...

Страница 17: ...e board provides a breadboard area with power and additional screw terminals to access the user designed circuitry EXP 16 and EXP 16 A expansion boards Sixteen channel multiplexer and signal condition...

Страница 18: ...S 1800 cable Shielded 18 inch cable with a 37 pin connector on each end allows you to connect a DAS 800 Series board to an STA 08 STA 08PGA EXP 16 EXP 16 A or EXP GP C8 MB1 cable Cable with a 37 pin...

Страница 19: ...is chapter describes the following features of DAS 800 Series boards Analog input features Digital I O features Counter timer I O features Interrupts Power A functional block diagram of a DAS 800 Seri...

Страница 20: ...rsion frequency of 40 kHz Clk 0 Gate 0 Out 0 Clk 2 Gate 2 Out 2 Clk 1 Gate 1 Out 1 Timer Counter MUX Clk 0 FIFO INT_IN XCLK IP1 TRIG Digital I O Analog In Ch 0 Channel Select Scan Logic Control Logic...

Страница 21: ...de common mode noise rejection Differential A differential input configuration is appropriate if you are measuring low level signals if high source resistances greater than 100 exist or if common mode...

Страница 22: ...on a DAS 801 or DAS 802 board you can specify one of five bipolar and four unipolar analog input ranges Table 2 1 lists the gains supported by DAS 800 Series boards and the analog input voltage range...

Страница 23: ...e channel You use software to specify a single channel and initiate a conversion Automatic channel scanning You use software to specify the first and last channels in a range of contiguous on board ch...

Страница 24: ...ies to the first channel DriverLINX uses the Stop Channel s gain for all the other analog channels in the range If the Start Channel is greater than the Stop Channel the channel sequence is Start Chan...

Страница 25: ...aisy chain is associated with on board channel 0 the next expansion board is associated with on board channel 1 and so on You specify the associated on board channel by setting a jumper on each expans...

Страница 26: ...using a software conversion clock the host computer issues a command to initiate a conversion The host polls the board to determine if the conversion is complete When the conversion is complete the ho...

Страница 27: ...r timer circuitry When C T2 counts down to 0 C T1 decrements by 1 C T2 is reloaded with its count value and begins counting down again Each time C T2 counts down to 0 C T1 decrements by 1 Each time bo...

Страница 28: ...a conversion is initiated Note If you are using a hardware external clock source you cannot use the INT_IN XCLK pin pin 24 to generate interrupts Figure 2 3 illustrates how conversions are initiated...

Страница 29: ...number of factors including your computer the operating system environment whether you are using expansion boards the gains of the channels and software issues You can synchronize conversions on mult...

Страница 30: ...in I O connector The trigger event occurs when the board detects a rising edge on IP1 TRIG The actual point at which conversions begin depends on whether you are using an internal or external clock so...

Страница 31: ...r Figure 2 4 Initiating Conversions with a Hardware Trigger Hardware Trigger Trigger event occurs Conversions begin when using an external clock source Conversions begin when using an internal clock s...

Страница 32: ...software starts the analog input operation the board waits until the gate signal goes high before conversions begin When using the hardware gate the way conversions are synchronized depends on whether...

Страница 33: ...er and a digital trigger Refer to the Using DriverLINX With Your Hardware Keithley DAS 800 manual for more information Software starts the operation 1st conversion external clock Internal Clock Source...

Страница 34: ...ns under hardware control using an internal or external clock source the result of each conversion is automatically stored in the FIFO Note When you use software to initiate conversions the FIFO contr...

Страница 35: ...four digital output lines control the multiplexers on the expansion boards to determine the expansion board channel that is acquiring data in this case you cannot use the digital output lines for gene...

Страница 36: ...these counters externally wire the output of counter C T2 to the clock input of counter C T1 Notes For compatibility with the DAS 8 board on power up or system reset the DAS 800 board connects the cl...

Страница 37: ...remains high until the software writes a new count value Note that the output does not go high until n 1 clock pulses after the initial count is written where n indicates the loaded count A high gate...

Страница 38: ...cycles in duration where n indicates the loaded count After the counter reaches zero the output goes high and remains high until the clock pulse after the next trigger this makes the one shot pulse re...

Страница 39: ...then reloaded with the initial count value and the process repeats A high gate input enables counting a low gate input disables counting If the gate goes low during an output pulse the output is set h...

Страница 40: ...r is reloaded with the initial count value the output goes high again and the process repeats If the initial count is odd the output is high for n 1 2 counts and low for n 1 2 counts where n indicates...

Страница 41: ...output goes low for one clock pulse and then goes high again Note that the output does not go low until n 1 clock pulses after the initial count is written where n indicates the loaded count A high g...

Страница 42: ...ot go low until n 1 clock pulses after the trigger event occurs where n indicates the loaded count After the trigger event occurs the gate input has no effect on the output Writing a new value during...

Страница 43: ...ersion EOC interrupt An interrupt is generated when an A D conversion is complete The interrupt source is software selectable At power up or system reset the board assumes that the interrupt source is...

Страница 44: ...rdware and software the board generates an interrupt to the host driving the selected host interrupt line to an active state Power The analog circuitry on the DAS 800 board is powered by the 12 V of t...

Страница 45: ...ion for information on providing external power to EXP 16 EXP 16 A and EXP GP expansion boards Refer to the MB Series User s Guide for information on providing external power to MB Series backplanes T...

Страница 46: ...pers and with the items that are configurable on DAS 800 Series boards you can use Figure 3 1 as a quick reference for configuring a DAS 800 board and Figure 3 2 as a quick reference for configuring a...

Страница 47: ...1 DAS 800 Board J2 2 3 4 5 6 7 X IRQ LEVEL 1 2 3 4 5 6 7 O N BASE ADDRESS Switch block S1 Pin 1 Main I O connector Interrupt level X disabled Base Address On 0 Off 1 0000000 000H 0000001 008H 0000010...

Страница 48: ...l X disabled Base Address On 0 Off 1 0000000 000H 0000001 008H 0000010 010H 1100000 300H 1111111 3F8H 1 2 3 4 5 6 7 O N Switch block S1 8 Input Configuration On single ended Off differential Switch 1...

Страница 49: ...standard software package Shipped with DAS 800 Series boards Includes DriverLINX for Microsoft Windows and function libraries for writing application programs such as Microsoft Visual C Microsoft Visu...

Страница 50: ...iverLINX Test Panel Utility Installing the DAS 800 Series Standard Software Package Note Important Before you begin installing any hardware or software for the DAS 800 read the DriverLINX Installation...

Страница 51: ...ly the files you need for configuring your hardware and running third party data acquisition applications that require DriverLINX Install Interfaces This optional component installs the files and exam...

Страница 52: ...ng Install Adobe Acrobat 6 Open the manuals appropriate to the DAS 800 installation and read them before installing your DAS 800 board or configuring DriverLINX Installation and Configuration Using Dr...

Страница 53: ...urned OFF but grounded hold the wrapped board in one hand while placing your other hand firmly on a metal portion of the computer chassis this discharges any static electricity 3 Carefully remove the...

Страница 54: ...es to perform DAS 800 Series board operations Use of C T2 on the 8254 cascaded or normal Input range type unipolar or bipolar for a DAS 801 or DAS 802 board Input configuration single ended or differe...

Страница 55: ...rLINX under Windows 95 98 the Add New Hardware automatically selects an appropriate address To change the address refer to the Using DriverLINX with Your Hardware Keithley DAS 800 manual The base addr...

Страница 56: ...e appropriate base address Table 3 1 I O Address Map 000H to 3FFH Address Range Use 000H to 00FH 8237 DMA 1 020H to 021H 8259 PIC 1 040H to 043H 8253 timer 060H to 063H 8255 PPI XT 060H to 064H 8742 c...

Страница 57: ...l port 2F8H to 2FFH Serial port 300H to 31FH Prototype card 320H to 32FH Hard disk XT 378H to 37FH Parallel printer 380H to 38FH SDLC 3A0H to 3AFH SDLC 3B0H to 3BBH MDA 3BCH to 3BFH Parallel printer 3...

Страница 58: ...0 Series boards have 8 analog input channels On DAS 801 and DAS 802 models you can switch each channel to differential or single ended The switch setting affects only the connections for the channel D...

Страница 59: ...nce DAS 800 boards always use a single ended input configuration they do not contain an input configuration switch block Setting the Interrupt Level DAS 800 Series boards are shipped with interrupts d...

Страница 60: ...ting the Interrupt Level Table 3 2 lists interrupt levels commonly used by IBM PC XT AT and compatible computers Select an interrupt level that is not being used by another resource in your system inc...

Страница 61: ...DAS 802 require a single 1 2 slot 4 Loosen and remove the screw at the top of the blank adapter plate and then slide the plate up and out to remove 5 Insert and secure the board connector in the selec...

Страница 62: ...ndows as you normally would and select the Program Manager window Install DriverLINX if you have not previously done so 2 Either select the Learn DriverLINX icon created when you installed DriverLINX...

Страница 63: ...and Counter Timer Sections by first clicking on the appropriate radio button in the middle of the dialog box and then completing the group of dialog fields in the lower third of the dialog box Be sur...

Страница 64: ...AS 800 Series board are OFF Attaching Accessory and Expansion Boards You attach an STC 37 screw terminal connector STA 08 or STA 08PGA screw terminal accessory or EXP 16 EXP 16 A or EXP GP expansion b...

Страница 65: ...12 DIG COM Pin 11 OP4 Pin 10 OP3 Pin 9 OP2 Pin 8 OP1 Pin 7 OUT2 Pin 6 OUT1 Pin 5 CLK1 Pin 4 OUT0 Pin 3 CLK0 Pin 2 12 V Pin 1 Pin 37 IN0 Pin 36 IN1 Pin 35 IN2 Pin 34 IN3 Pin 33 IN4 Pin 32 IN5 Pin 31 I...

Страница 66: ...to a DAS 800 Series board The screw terminals are labeled from 1 to 37 and correspond directly to the functions of the pins on the main I O connector on the board For example since pin 24 is assigned...

Страница 67: ...erminal Accessory The screw terminals on the STA 08 screw terminal accessory allow you to connect field wiring to a DAS 800 board the screw terminals on the STA 08PGA screw terminal accessory allow yo...

Страница 68: ...expansion board provides up to 16 analog input channels labeled 0 to 15 The EXP 16 and EXP 16 A provide three screw terminals for each channel low level ground LL GND positive input CHn HI and negativ...

Страница 69: ...inals for each channel positive current excitation IEXC positive input SENSE negative voltage excitation P negative current excitation IEXC negative input SENSE and positive voltage excitation P To co...

Страница 70: ...ansion boards to provide up to 128 analog input channels You connect the first expansion board to the DAS 800 Series board by attaching one end of an S 1800 or C 1800 cable to the main I O connector o...

Страница 71: ...sion board in the chain to attach an STA 08 STA 08PGA screw terminal accessory This allows you to access the remaining functions of the DAS 800 Series board Figure 4 7 illustrates how to connect two E...

Страница 72: ...2 boards contain digital commons but no dedicated low level analog commons If you are using all eight analog input channels and all eight channels are configured for differential input you must connec...

Страница 73: ...input use the negative side of that channel as the low level analog common otherwise use the digital common Figure 4 9 illustrates how to connect an analog input signal to differential analog input ch...

Страница 74: ...ecommended that the sum of the resistors used be at least 5 000 times greater than the source output resistor Voltage Source Pin 37 Chan 0 Pin 19 Chan 0 DAS 801 Board Pin 12 Chan 7 Voltage Source Pin...

Страница 75: ...upt to the INT_IN XCLK pin pin 24 on the main I O connector Make sure that all digital signals are TTL level compatible Use the digital common as the return for all digital signals Caution To prevent...

Страница 76: ...Series boards Figure 4 10 illustrates how you can synchronize conversions on multiple boards using both an external clock source and a DAS 802 board Note that in both cases boards 1 2 and 3 must be c...

Страница 77: ...o a rate different from that of the master board Note that when you use C T0 to divide the rate of the master clock C T0 must be configured for counter timer mode 2 rate generator mode Refer to page 2...

Страница 78: ...ing the SST Signal Generator Output DC Level voltages using the Level Control The Analog I O Panel is useful for Testing the DAS 800 DriverLINX installation and configuration Verifying signal inputs t...

Страница 79: ...two analog output channels referred to as ChA and ChB Drag the channel selectors in the AO Channel Mapping section to map them to different channel numbers You can now select the Scope Meter SST and L...

Страница 80: ...he following sections provide the information you need to calibrate DAS 800 Series boards Equipment Required To calibrate a DAS 800 Series board you need either A digital voltmeter accurate to 6 1 2 d...

Страница 81: ...0 board contains three potentiometers R1 full scale R2 offset and R3 10 V reference The DAS 801 and DAS 802 boards contain four potentiometers R1 offset R2 full scale R5 unipolar 0 and R6 high gain 0...

Страница 82: ...6 3 Figure 6 1 Potentiometers DAS 800 Figure 6 2 Potentiometers DAS 801 DAS 802 R1 R2 R3 Full scale Offset 10 V reference R1 R2 R5 Offset Full scale High gain 0 R6 Unipolar 0 TP1 TP2...

Страница 83: ...eries manuals Before calibration specify the following parameters in the setup panel to get the correct instructions Logical Device Board s device number model and address Accessory Connection method...

Страница 84: ...assist in predicting and troubleshooting hardware and software problems DriverLINX uses the event log to report problems during driver loading or unexpected system errors The event log can assist in...

Страница 85: ...dware does not match configuration Operating mode range switch or jumper setting does not match selection s made in the DriverLINX Device Configuration dialog box Problem Isolation If you encounter a...

Страница 86: ...0 Series board check the other slots you are using 7 If the accessory slots are functional check the I O hookups Connect the accessory and expansion boards one at a time and check operation 8 If opera...

Страница 87: ...ort Table 7 1 Troubleshooting Information Symptom Possible Cause Possible Solution Board does not respond Base address is unacceptable Make sure that the base address specified in the configuration fi...

Страница 88: ...in the accessory slot Check installation The board is damaged Contact the Keithley Data Acquisition Applications Engineering Department refer to page 7 7 Intermittent operation Vibrations or loose co...

Страница 89: ...ive common mode voltages are not present Another system resource is using one of the eight memory locations starting at the specified base address Reconfigure the base address of the DAS 800 Series bo...

Страница 90: ...ng information available before you call DAS 800 board Model configuration Serial Number Revision Code Base address setting Interrupt level setting Number of channels Input S E or Diff Mode uni or bip...

Страница 91: ...the problem or its symptoms The RMA number on the outside of the package Repackage the equipment using the original anti static wrapping if possible and handle it with ground protection Ship the equi...

Страница 92: ...8 8 8 Full scale range gain 5 V 1 10 V 0 5 5 V 1 500 mV 10 50 mV 100 10 mV 500 0 to 10 V 1 0 to 1 V 10 0 to 100 mV 100 0 to 20 mV 500 10 V 0 5 5 V 1 2 5V 2 1 25 V 4 625 mV 8 0 to 10 V 1 0 to 5 V 2 0 t...

Страница 93: ...x Temperature coefficient of gain drift 45 ppm C maximum 50 ppm C maximum 50 ppm C maximum Temperature coefficient of zero drift3 100 V C maximum 1 100 gain V C maximum 1 100 gain V C maximum Common m...

Страница 94: ...re internal time base external clock falling edge software selectable Software internal time base external clock falling edge software selectable Maximum conversion frequency 40 kHz 40 kHz 40 kHz Gati...

Страница 95: ...r of outputs 4 OP1 to OP4 4 OP1 to OP4 4 OP1 to OP4 Input output levels TTL compatible TTL compatible TTL compatible Output high voltage 2 7 V minimum at 0 4 mA 2 7 V minimum at 0 4 mA 2 7 V minimum a...

Страница 96: ...1 to 0 Negative count changes on transition from 1 to 0 Negative count changes on transition from 1 to 0 Clock pulse width 50 ns high min 50 ns low min 50 ns high min 50 ns low min 50 ns high min 50 n...

Страница 97: ...ensing 0 to 90 noncondensing 0 to 90 noncondensing Dimensions 5 0 x 4 25 x 0 75 12 7 cm x 10 8 cm x 1 9 cm 6 0 x 4 25 x 0 75 15 2 cm x 10 8 cm x 1 9 cm 6 0 x 4 25 x 0 75 15 2 cm x 10 8 cm x 1 9 cm Not...

Страница 98: ...onnector and its pin assignments on a DAS 800 board Table B 1 contains a more detailed description of the pins Figure B 2 shows the main I O connector and its pin assignments on a DAS 801 or DAS 802 b...

Страница 99: ...11 OP4 Pin 10 OP3 Pin 9 OP2 Pin 8 OP1 Pin 7 OUT2 Pin 6 OUT1 Pin 5 CLK1 Pin 4 OUT0 Pin 3 CLK0 Pin 2 12 V Pin 1 Pin 37 IN0 Pin 36 IN1 Pin 35 IN2 Pin 34 IN3 Pin 33 IN4 Pin 32 IN5 Pin 31 IN6 Pin 30 IN7 P...

Страница 100: ...9 OP3 Digital output bit 2 10 OP4 Digital output bit 3 11 DIG COM Digital common 12 LL COM Low level common 13 LL COM Low level common 14 LL COM Low level common 15 LL COM Low level common 16 LL COM...

Страница 101: ...power from host computer 30 IN7 Analog input channel 7 31 IN6 Analog input channel 6 32 IN5 Analog input channel 5 33 IN4 Analog input channel 4 34 IN3 Analog input channel 3 35 IN2 Analog input chann...

Страница 102: ...4 Pin 10 OP3 Pin 9 OP2 Pin 8 OP1 Pin 7 OUT2 Pin 6 OUT1 CCLK Pin 5 CLK1 Pin 4 OUT0 Pin 3 CLK0 Pin 2 12 V Pin 1 Pin 37 IN0 Pin 36 IN1 Pin 35 IN2 Pin 34 IN3 Pin 33 IN4 Pin 32 IN5 Pin 31 IN6 Pin 30 IN7 Pi...

Страница 103: ...utput bit 1 9 OP3 Digital output bit 2 10 OP4 Digital output bit 3 11 DIG COM Digital common 12 IN7 Analog input channel 7 13 IN6 Analog input channel 6 14 IN5 Analog input channel 5 15 IN4 Analog inp...

Страница 104: ...r from host computer 30 IN7 Analog input channel 7 31 IN6 Analog input channel 6 32 IN5 Analog input channel 5 33 IN4 Analog input channel 4 34 IN3 Analog input channel 3 35 IN2 Analog input channel 2...

Страница 105: ...ansion boards 4 1 EXP 16 and EXP 16 A 4 5 EXP GP 4 6 Attaching screw terminal accessories 4 1 STA 08 STA 08PGA 4 4 STC 37 4 3 Attaching signals see Connecting signals Automatic channel scanning 2 5 B...

Страница 106: ...tor main I O see Main I O connector Conversion clocks 2 8 Conversion frequency 2 2 2 16 Conversion rate see Conversion frequency Converter see Analog to digital converter Counter timer modes 2 19 see...

Страница 107: ...expansion board configuring in DriverLINX 3 13 expansion board configuration DriverLINX dialog for expansion channels 2 6 Expansion boards 2 7 attaching 4 1 daisy chaining 4 7 EXP 16 and EXP 16 A 1 4...

Страница 108: ...tor see Main I O connector Jumper 3 15 L LabVIEW function libraries for writing application programs in 1 2 learn DriverLINX program in DriverLINX description of 1 3 used to configure DAS 800 3 17 log...

Страница 109: ...STA SCM8 1 5 STC 37 1 4 Setting base address 3 10 input configuration 3 13 interrupt level 3 14 Signal conditioning 1 4 2 7 Single channel selection 2 5 Single ended inputs connecting 4 9 Single ended...

Страница 110: ...log programming 2 15 digital programming 2 15 Triggers 2 12 Troubleshooting 7 1 Turbo Pascal function libraries for writing application programs in 1 2 U Unipolar input range type 2 4 Unpacking the bo...

Страница 111: ...e 65 82110 Germering 089 84 93 07 40 Fax 089 84 93 07 34 GREAT BRITAIN Unit 2 Commerce Park Brunel Road Theale Berkshire RG7 4AB 0118 929 7500 Fax 0118 929 7519 INDIA Flat 2B Willocrissa 14 Rest House...