RS-232 Commands
Refer to Table 3-2 for a list of commands included in the command set.
NOTE
ACSII characters not included in the command set are ignored.
Table 3-2
Command set
Function
ASCII Command Description
Open the relay
O or o
After the open command is sent, a 25msec delay
occurs which allows the relay to settle. After the
relay settles, an ASCII ‘1’ is sent over the RS-232
indicating settling time has run out.
Close the relay
C or c
After the closed command is sent, a 25msec delay
occurs which allows the relay to settle. After the
relay settles, an ASCII ‘1’ is sent over the RS-232
indicating settling time has run out.
Query relay state
S or s
Upon receiving this command, if the relay is closed
an ASCII 1 is returned; if the relay is open, an ASCII
0 is returned.
Firmware
information
I or i
Firmware information can be retrieved over the
RS-232 using the ‘I’ command. After the command
is sent, ‘Keithley Instruments Inc., Model 7999-4,
rev A01’ (or similar due to firmware revision level)
will be returned over the RS-232.