Model 7701
11
Figure 7
Connecting ribbon cable assembly
Solder cup cable connections
Make all connections to D-shell male solder cup connectors using the correct wire size up to 20 AWG.
Terminal identification for the 50-pin D-shell connector is provided in
. Terminal
identification for the 25-pin D-shell connector is provided in
and
. Make sure to add sup-
plementary insulation around the harness for voltages above 42V peak (
WARNING
All solder cup wiring must be rated for the maximum voltage in the system. For
example, if 150V is applied to the front terminals of the DMM, All matrix module
wiring must be rated for 150V.
A connector cover must be installed on an unused D-shell connector. If the connec-
tor is left open, an electrical shock hazard may be present.
Figure 8
Connecting solder cup cable assembly
Install plastic cover
over unused connector
using two #4-40 screws.
KEITHLEY
CONNECTOR COVER
Warning:
When using ribbon cable,
DO NOT exceed 42 volts
anywhere in the test system
or at the front panel inputs
of the Model 2700/2750.
Supplementary
Insulation
Install plastic cover
over unused connector
using two #4-40 screws.
KEITHLEY
CONNECTOR COVER