Section 7: Command reference
Models 707B and 708B Switching Matrix Reference Manual
7-62
707B-901-01 Rev. A / August 2010
digio.trigger[N].reset()
This function resets trigger values to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
digio.trigger[N].reset()
N
Digital I/O trigger line (1 to 14)
Details
This function resets the following attributes to factory default settings:
•
digio.triggger[N].mode
•
digio.trigger[N].pulsewidth
•
digio.trigger[N].stimulus
It also clears
digio.trigger[N].overrun
.
Also see
(on page 7-58)
(on page 7-60)
(on page 7-60)
digio.trigger[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
System reset
Digital I/O trigger reset
Create configuration script
0
Usage
triggerStimulus = digio.trigger[N].stimulus
digio.trigger[N].stimulus = triggerStimulus
triggerStimulus
The event identifier for the triggering event
N
Digital I/O trigger line (1 to 14)
Details
Set this attribute to zero (0) to disable the automatic trigger output.
Trigger stimulus for a digital I/O line
may be set to one of the existing trigger event IDs, described in the following
table.
Do not use the stimulus attribute for generating output triggers. Use
digio.trigger[N].assert()
instead.