Service Information
5-6
Figure 5-4
Channel to channel testing
Refer to Figure 5-4 and perform the following steps to test
channel to channel isolation:
1. Turn off the Model 7001/7002 if it is on.
2. Connect the Model 6517A to channels 1 and 2 as shown
in Figure 5-4. Make sure the voltage source is off.
3. Install the Model 7036 in slot 1 (CARD 1) of the
Model 7001/7002 if it is not already installed.
4. On the Model 6517A, select the ohms function and
choose the 2G
Ω
range.
5. On the Model 6517A, set the voltage source for +42V.
Make sure the voltage source is off.
6. Place the Model 6517A in the R measurement function.
7. Turn on the Model 7001/7002, and program it to close
channels 1!1 and 1!2 (channels 1 and 2).
8. On the Model 6517A, turn on the output of the voltage
source.
9. After allowing the reading on the Model 6517A to settle,
verify that it is >1G
Ω
. This measurement is the channel
to channel leakage resistance (isolation) between chan-
nels 1 and 2.
10. Turn off the Model 6517A voltage source.
11. Turn off the Model 7001/7002.
12. Disconnect the Model 6517A from channels 1 and 2,
and, in a similar manner, reconnect it to channels 2 and
3 (electrometer high to channel 2, and voltage source
high to channel 3).
13. Install the card in slot 1 (CARD 1) of the
Model 7001/7002 if it is not already installed.
14. Turn on the Model 7001/7002 and program it to close
channels 1!2 and 1!3.
15. On the Model 6517A, turn on the voltage source.
16. After allowing the reading on the Model 6517A to settle,
verify that it is >1G
Ω
. This is the isolation between
channels 2 and 3.
17. Using Table 5-2 as a guide, repeat the basic procedure of
steps 11 through 16 for the rest of the path pairs (starting
with test 3).
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
!
115V
!
90-110V
105-125V
180-220V
210-250V
Banana to
Banana Cable
7036
IN
OUT
Each Adjacent Channel Pair:
IN
OUT
HI
Model 6517A
Model 7025
Unterminated
Triax Cable
(Red
Ground Link
Removed
Unterminated
Banana Cables
Source V and
Measure V/I
HI