Multiplexer Configuration
2-6
Separate switching systems
Two single-card systems can be configured by removing the
backplane jumpers from one of the cards. The two cards will
be controlled by the same mainframe, but they will be elec-
trically isolated from each other. Figure 2-10 shows an exam-
ple using two Model 7021 cards.
Multiplexer input expansion
You can double the number of multiplexer inputs by install-
ing two “as shipped” Model 7021s in the Model 7001/7002
mainframe. By leaving the backplane jumpers installed, the
multiplex banks of the card installed in slot 1 (CARD 1) are
automatically connected to the multiplex banks of the card
installed in slot 2 (CARD 2) through the analog backplane.
Figure 2-10
Two separate multiplexer systems
7021
Card 1
1x12 and 1x18 Multiplexers
7021
Card 2
Backplane
Jumpers
Removed
1x12 and 1x18 Multiplexers
Bank B
Bank A
7001/7002
Analog Backplane
4
9
14
19
24
29
5
10
15
20
25
30
1
26
6
11
16
21
2
27
7
12
17
22
3
8
13
18
23
28
4
9
14
19
24
29
5
10
15
20
25
30
1
26
6
11
16
21
2
27
7
12
17
22
3
8
13
18
23
28
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Содержание 7021
Страница 63: ...Operation 5 16 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 86: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 87: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 88: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 89: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Страница 90: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...