3-12
Remote Operation
Model 6517A Getting Started Manual
Function configuration
[:SENSe[1]]
SENSe
Subsystem
:FUNCtion <name>
Select measurement function:
<name>
=
'VOLTage[:DC]'
Volts function
'CURRent[:DC]'
Amps function
'RESistance'
Ohms function
'CHARge'
Coulombs
function
:VOLTage[:DC]
Path to configure volts:
:NPLC <n>
Set speed: 0.01 to 10 PLC.
:AUTO <b>
Enable or disable auto NPLC.
:RANGe
Path to configure measurement range:
[:UPPer] <n>
Specify expected reading to select range.
:AUTO <n>
Enable or disable auto range.
:REFerence <n>
Specify reference (rel): -210 to +210.
:STATe <b>
Enable or disable reference (rel).
:ACQuire
Use input signal as reference (rel).
:DIGits <n>
Specify measurement resolution.
:AUTO <b>
Enable or disable auto resolution.
AVERage
Path to control digital filter:
:TYPE <name>
Select filter (NONE, SCALar or ADVanced).
:TCONtrol <name>
Select averaging method (REPeat, or
MOVing).
:COUNt <n>
Specify filter count (1 to 100).
:ADVanced
Path to configure advanced filter:
:NTOLerance <n>
Specify noise tolerance (0 to 100%).
[:STATe] <b>
Enable or disable digital filter.
:MEDian
Path to control median filter:
:RANK < n>
Specify rank for filter (1 to 5).
[:STATe] <b>
Enable or disable median filter.
:GUARd <b>
Enable or disable guard.
:CURRent:DC
Path to configure Amps:
(Volts commands)
The commands for Volts (except :GUARd)
apply.
:DAMPing <b>
Enable or disable damping.