1-12
Figure 1-5
Connections for ohms verification (2M
Ω
–200M
Ω
ranges)
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
INPUT
250V PEAK
!
LINE RATING
50-60HZ
50VA MAX
AC ONLY
LINE FUSE
SLOWBLOW
1/2A 90-125V
1/4A 180-250V
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
DIGITAL
I/O
TRIG LINK
115V
LO
HI
RS232
!
MADE
IN
U.S.A.
PREAMP OUT
250 PEAK
COMMON
2V ANALOG
OUTPUT
TEMP
TYPE K
HUMIDITY
OPTION SLOT
V SOURCE
EXT TRIG
IN
MTR COMP
OUT
INTERLOCK
SELECTED
LINE VOLTAGE
90-110V
105-125V
180-220V
210-250V
1010V
PEAK
750V
PEAK
Triax Cable
Model 6517 Electrometer
HI
Resistance Calibrator
Voltage Source HI
Table 1-9
Limits for ohms verification
Model
6517A
range
Nominal
resistance
1
Reading limits
2
(1 year, 18°–28°C)
2M
Ω
20M
Ω
200M
Ω
2G
Ω
20G
Ω
200G
Ω
2T
Ω
20T
Ω
200T
Ω
1.9M
Ω
19M
Ω
100M
Ω
1G
Ω
10G
Ω
100G
Ω
1T
Ω
10T
Ω
100T
Ω
___________ to __________ M
Ω
___________ to __________ M
Ω
___________ to __________ M
Ω
___________ to __________ G
Ω
___________ to __________ G
Ω
___________ to __________ G
Ω
___________ to __________ T
Ω
___________ to __________ T
Ω
___________ to __________ T
Ω
1
Resistance calibrator used for 2M
Ω
-200M
Ω
ranges. Model 5156 Calibration
Standard used for 2G
Ω
-200G
Ω
ranges. Characterized resistance standards
used for 2T
Ω
-200T
Ω
ranges. Standards must be characterized to uncertainty
at least four times better than equivalent Model 6517A specifications and
mounted in shielded test box. (See Figure 1-8.)
2
Reading limits calculated from actual resistance value and Model 6517A
specifications. See text.