Maintenance
7-13
Table 7-8
Power supply checks
Step
Item/component
Required condition
Remarks
1
2
3
4
5
6
7
8
9
S102 Line Switch
F101 Line Fuse
Line power
+5V Digital Supply
±210V Supplies
+5V Analog Supply
-9.1V Supply
±5V Bootstrap Supplies
±24 Supplies
Set to 115 or 230 as required.
Check for continuity.
Plugged into live receptacle; power on.
5V, ±5%
+240V, -240V, ±15%
+5V, ±5%
-9.1V, ±5%
+5V, -5V, ±5%
+27.5V, -27.5V, ±15%
Line voltage selection.
Remove fuse to check.
Referenced to digital common.
Referenced to analog common.
Referenced to analog common.
Referenced to analog common.
Referenced to bootstrap common.
Referenced to preamp out.
Table 7-9
Relay configuration
Relay
Function
Range
K309
K310
K308
K311
K312
K307
K303
K304
K305
K306
K301
K302
Volts
200mV
2V
20V
200V
X
X
X
X
Ohms
2k
Ω
20k
Ω
200k
Ω
2M
Ω
20M
Ω
200M
Ω
2G
Ω
20G
Ω
200G
Ω
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Amps
2pA
20pA
200pA
2nA
20nA
200nA
2µA
20µA
200µA
2mA
20mA
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Coulombs
200pC
2nC
20nC
X
X
X
X
X
X
X
X
X
External
Feedback
200mV
2V
20V
X
X
X
X
X
X
Zero
Check
*
*
*
*
*
X
*
*
*