Model 6485 Picoammeter Instruction Manual
Remote Operation
9-9
GET (group execute trigger)
GET is a GPIB trigger that is used as an event to control operation. Model 6485 reacts to
this trigger if it is the programmed control source. The control source is programmed from
the SCPI TRIGger subsystem.
SPE, SPD (serial polling)
Use the serial polling sequence to obtain Model 6485 serial poll byte. The serial poll byte
contains important information about internal functions. Generally, the serial polling
sequence is used by the controller to determine which of several instruments has requested
service with the SRQ line. However, the serial polling sequence may be performed at any
time to obtain the status byte from Model 6485.
Front panel GPIB operation
The following paragraphs describe aspects of the front panel that are part of GPIB opera-
tion, including messages, status indicators, and the LOCAL key.
Error and status messages
See
for a list of error and status messages associated with IEEE-488 program-
ming. The instrument can be programmed to generate an SRQ, and command queries can
be performed to check for specific error conditions.
GPIB status indicators
The REM (remote), TALK (talk), LSTN (listen), and SRQ (service request) annunciators
show the GPIB bus status. Each of these indicators is described below.
•
REM — This indicator shows when the instrument is in the remote state. REM
does not necessarily indicate the state of the REM line, as the instrument must be
addressed to listen with REM true before the REM indicator turns on. When the
instrument is in remote, all front panel keys, except for the LOCAL key, are locked
out. When REM is turned off, the instrument is in the local state, and front panel
operation is restored.
•
TALK — This indicator is on when the instrument is in the talker active state. Place
the unit in the talk state by addressing it to talk with the correct MTA (My Talk
Address) command. TALK is off when the unit is in the talker idle state. Place the
unit in the talker idle state by sending a UNT (Untalk) command, addressing it to
listen, or sending the IFC (Interface Clear) command.
•
LSTN — This indicator is on when Model 6485 is in the listener active state, which
is activated by addressing the instrument to listen with the correct MLA (My Listen
Address) command. LSTN is off when the unit is in the listener idle state. Place the
unit in the listener idle state by sending UNL (Unlisten), addressing it to talk, or
sending the IFC (Interface Clear) command over the bus.
Содержание 6485
Страница 1: ...Model 6485Picoammeter Instruction Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E ...
Страница 20: ......
Страница 24: ......
Страница 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Страница 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Страница 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Страница 159: ...11 CommonCommands ...
Страница 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Страница 165: ...12 SCPISignalOriented MeasurementCommands ...
Страница 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Страница 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Страница 219: ...A Specifications ...
Страница 221: ...B StatusandErrorMessages ...
Страница 227: ...C GeneralMeasurement Considerations ...
Страница 233: ...D DDCEmulationCommands ...
Страница 247: ...E ExamplePrograms ...
Страница 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Страница 251: ...F IEEE 488BusOverview ...
Страница 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Страница 267: ...G IEEE 488andSCPI ConformanceInformation ...
Страница 311: ......