SCPI Command Reference
17-77
:POINts?
:SOURce[1]:LIST:CURRent:POINts?
Query length of I-Source list
:SOURce[1]:LIST:VOLTage:POINts?
Query length of V-Source list
Description
This command is used to determine the length of the speci
fi
ed source list.
The response message indicates the number of source values in the list.
Con
fi
gure memory sweep
A memory sweep lets you perform a sweep using setups stored in up to 100 memory loca-
tions. This allows you to sweep using multiple source-measure operations and math expres-
sions. See
CALCulate1
.
NOTE
In order to run a memory sweep, the memory function must be selected, and the trig-
ger count must be the same as the number of memory points in the sweep. Use the
:FUNCtion:MODE command to select the MEMory function. See “Select function
mode.” Arm count (ARM:COUNt) and trigger count (TRIGger:COUNt) are set from
the Trigger Subsystem.
When memory is initialized (:SYSTem:MEMory:INITialize), all 100 memory loca-
tions for a memory sweep are initialized to the present setup configuration of the
SourceMeter with CALC 1 disabled. User-defined math expressions are replaced
with the “Power” math expression.
Error 809 “Source memory location revised” occurs when a memory sweep refer-
ences an expression that no longer exists. The memory sweep is revised to disable
CALC1.
In order to execute a memory sweep, the product of the arm count and trigger count
should be at least the same as the number of points in the sweep.
Use the :FUNCtion:MODE command to select the MEMory function. See
Select function
mode
. Arm count (ARM:COUNt) and trigger count (TRIGger:COUNt) are set from the Trigger
Subsystem.
:SAVE <NRf>
:SOURce[1]:MEMory:SAVE <NRf>
Save setup in speci
fi
ed memory location
Parameters
<NRf> = 1 to 100
Specify memory location
Description
This command is used to save the present instrument setup in a memory
location. Up to 100 setups can be saved. The following settings are saved in
each source memory location:
SENSe[1]:CURRent:NPLCycles
SENSe[1]:RESistance:NPLCycles
SENSe[1]:VOLTage:NPLCycles
SENSe[1]:FUNCtion:CONCurrent
SENSe[1]:FUNCtion:ON
SENSe[1]:FUNCtion:OFF
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......