17-44
SCPI Command Reference
:STATe <b>
:DISPlay[:WINDow[1]]:TEXT:STATe <b>
Control message; top display
:DISPlay:WINDow2:TEXT:STATe <b>
Control message; bottom display
Parameters
<b> =
0 or OFF
Disable text message for speci
fi
ed display
1 or ON
Enable text message for speci
fi
ed display
Query
:STATe?
Query state of message mode for speci
fi
ed
display
Description
These commands enable and disable the text message modes. When
enabled, a de
fi
ned message is displayed. When disabled, the message is
removed from the display.
GPIB Operation
—
A user de
fi
ned text message remains displayed only as
long as the instrument is in remote. Taking the instrument out of remote (by
pressing the LOCAL key or sending LOCAL 27) cancels the message and
disables the text message mode.
RS-232 Operation
—
A user de
fi
ned test message can be cancelled by
sending the :SYSTem:LOCal command or pressing the LOCAL key.
FORMat subsystem
The commands for this subsystem are used to select the data format for transferring instru-
ment readings over the bus. These commands are summarized in Table 17-3.
Data format
[:DATA] <type>[,length]
:FORMat[:DATA] <type>[,<length>]
Select data format
Parameters
<type>[,<length>] = ASCii
ASCII format
REAL,32
IEEE754 single precision format
SREal
IEEE754 single precision format
NOTE
<length> is not used for the ASCii or SREal parameters. It is optional for the REAL
parameter. If you do not use <length> with the REAL parameter, the <length>
defaults to 32 (single precision format).
Query
[:DATA]?
Query data format
Description
This command is used to select the data format for transferring readings
over the bus. Only the ASCII format is allowed over the RS-232 interface.
This command only affects the output of READ?, FETCh?, MEASure?,
TRACe:DATA?, CALC1:DATA? and CALC2:DATA? over the GPIB. All
other queries are returned in the ASCII format.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......