Status Structure
14-17
Event registers
As Figure 14-1 shows, each status register set has an event register. When an event occurs,
the appropriate event register bit sets to 1. The bit remains latched to 1 until the register is reset.
Reading an event register clears the bits of that register. *CLS resets all four event registers.
The commands to read the event registers are listed in Table 14-6. For details on reading reg-
isters, see
Reading registers
.
Event enable registers
As Figure 14-1 shows, each status register set has an enable register. Each event register bit
is logically ANDed (&) to a corresponding enable bit of an enable register. Therefore, when an
event bit is set and the corresponding enable bit is set (as programmed by the user), the output
(summary) of the register will set to 1, which in turn sets the summary bit of the Status Byte
Register.
The commands to program and read the event enable registers are listed in Table 14-7. For
details on programming and reading registers, see
Programming enable registers
and
Reading
registers
.
NOTE
The bits of any enable register can be reset to 0 by sending the 0 parameter value
with the appropriate enable command (i.e. STATus:OPERation:ENABle 0).
Table 14-5
Condition register commands
Command
Description
:STATus:OPERation:CONDition?
:STATus:MEASurement:CONDition?
:STATus:QUEStionable:CONDition?
Read Operation Condition Register.
Read Measurement Condition Register.
Read Questionable Condition Register.
Table 14-6
Event register commands
Command
Description
Default
*ESR?
:STATus:OPERation:[:EVENt]?
:STATus:MEASurement:[:EVENt]?
:STATus:QUEStionable:[:EVENt]?
Read Standard Event Status Register.
Read Operation Event Register.
Read Measurement Event Register.
Read Questionable Event Register.
Note
Note: Power-up and *CLS resets all bits of all event registers to 0. STATus:PRESet has no effect.
Содержание 6430
Страница 26: ......
Страница 32: ......
Страница 78: ...2 14 Connections ...
Страница 98: ...3 20 Basic Source Measure Operation ...
Страница 138: ...5 30 Source Measure Concepts ...
Страница 156: ...6 18 Range Digits Speed and Filters ...
Страница 168: ...7 12 Relative and Math ...
Страница 176: ...8 8 Data Store ...
Страница 202: ...9 26 Sweep Operation ...
Страница 248: ...11 22 Limit Testing ...
Страница 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Страница 418: ...17 108 SCPI Command Reference ...
Страница 450: ...18 32 Performance Verification ...
Страница 477: ...A Specifications ...
Страница 489: ...B StatusandErrorMessages ...
Страница 498: ...B 10 Status and Error Messages ...
Страница 499: ...C DataFlow ...
Страница 503: ...D IEEE 488BusOverview ...
Страница 518: ...D 16 IEEE 488 Bus Overview ...
Страница 519: ...E IEEE 488andSCPI ConformanceInformation ...
Страница 523: ...F MeasurementConsiderations ...
Страница 539: ...G GPIB488 1Protocol ...
Страница 557: ......