4-14
Sweeps
Model 6220/6221 Reference Manual
Return to
Remote sweep operation
Procedures for programming and running a sweep for each of the three sweep types
are given on the following pages. Each of these procedures includes commands for a
typical sweep example.
summarizes parameters for each of these exam-
ples. See
“SCPI commands — sweeps,” on page 4-18
for details on sweep com-
mands.
NOTE
User setups cannot be saved or recalled while
sweep is armed or running. The *SAV and *RCL
commands will generate error +413 Not allowed
with mode armed.
Table 4-3
Sweep example parameters
Sweep type
Parameters for sweep examples
Linear staircase sweep (
Start current: 1mA
Stop current: 10mA
Step current: 1mA
Delay: 1s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 100
μ
A
Compliance: 10V
Logarithmic staircase sweep (
)
Start current: 1mA
Stop current: 10mA
# points: 5
Delay: 2s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 10
μ
A
Compliance: 5V
Custom (list) sweep (
)
# points: 5
Points: 3mA, 1mA, 4mA, 5mA, 2mA
Delay: 3s, 1s, 4s, 5s, 2s
Compliance: 3V, 1V, 4V, 5V, 2V
Source range: auto
Compliance abort: off
Sweep count: 1
Bias current: 50
μ
A
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Содержание 6220
Страница 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 403: ...A Specifications Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Страница 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...