Keithley 4288-10 Скачать руководство пользователя страница 3

 

Model 4288-10 Rear Support Rack-Mount Kit Assembly and Mounting Instructions 

 

 

PA-953 Rev. E / July 2018 

 

Prepare the Model 3706A for rack mount installation 

Before installing the instrument, disconnect all external power from the equipment and 
disconnect the line cord. Failure to disconnect all power may expose you to hazardous 
voltages, which, if contacted, could cause personal injury or death. 

To prepare the instrument for mounting: 

1.  Verify that power is disconnected from the instrument. 

2.  Place the instrument upside-down. 

3.  From the four foot assemblies on the bottom of the instrument, remove the rubber centers. 

4.  Use a Phillips-head screwdriver to remove the four foot assemblies. Retain the rubber foot   

assemblies for future use. 

5.  Place the instrument right-side up. 

6.  Determine the placement of the rear support brackets (3706-354), using the measurement that   

you took when preparing the rack. Make sure the flanges on the rear support brackets are oriented   
to the outside of the instrument as shown in the following figure. 

7.  Use four of the #8-32 screws with washers to loosely attach the two rear support brackets to the 

Model 3706A. You will use two screws and washers on each side. Do not tighten the screws fully   
at this time. 

Figure 2: Rear support bracket installation 

 

 

Model 3706A 

Flanges of the rear support brackets (part number 3706-354) 

#8-32 screws 

#8 x 3/4 washer 

 

Содержание 4288-10

Страница 1: ...rack mount hardware that is supplied with the Model 3706A The supplied kit provides support for the front of the instrument The Model 4288 10 provides the support for the rear The Model 4288 10 Rear...

Страница 2: ...these sensors must be in stable ambient conditions not subjected to hot spots Prepare the rack for Model 3706A installation To prepare the rack for Model 3706A installation 1 If possible remove the c...

Страница 3: ...emove the rubber centers 4 Use a Phillips head screwdriver to remove the four foot assemblies Retain the rubber foot assemblies for future use 5 Place the instrument right side up 6 Determine the plac...

Страница 4: ...the remaining four 10 32 screws 4 Tighten the 10 32 screws to secure the instrument to the rack 5 Tighten the 8 32 screws that secure the rear support brackets to the Model 3706A Install the support b...

Страница 5: ...ation in this documentation please contact your local Keithley Instruments office sales partner or distributor You can also call the corporate headquarters of Keithley Instruments toll free inside the...

Страница 6: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Страница 7: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Отзывы: