Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Section 5: KPulse (for Keithley Pulse Cards)
4200A-PMU-900-01 Rev. A December 2020
5-7
Select
Reset All
to return the pulse card to the Standard Pulse waveform type and its default
settings. It also updates the previewer.
To configure the Channel 1 and Channel 2 settings:
1. Configure the channel settings as needed. Note that the Pulse Count field is only available if the
Output Mode is set to Burst Mode.
2. At the bottom of the window, select
Enable Channel 1
or
Enable Channel 2
. A channel must be
enabled to preview its waveform and turn on the output for the channel.
3. Select
Edit Segarb
.
4. For each segment, enter the
Start
voltage,
Stop
voltage,
Time
in seconds, the TTL output
Trig
level (
0
= low;
1
= high), and the state of the
SSR
(solid-state relay,
0
= open,
1
= closed).
5. Select
OK
.
6. Select the
Output On
button to turn on all enabled channels for all pulse cards installed on the
4200A-SCS. Output Off turns red.
7. To turn off the output, select
Output Off
.
Exporting Segment Arb waveform files
After configuring a Segment Arb
®
waveform in KPulse, you can save it as a
.ksf
file. The Segment
Arb
.ksf
files should be exported into the
SarbFiles
folder at the following location:
C:\s4200\kiuser\KPulse\SarbFiles
To export a Segment Arb waveform file:
1. Select
Tools
.
2. Select
Export Segment Arb
to open the Segment Arb Export dialog box.
Figure 104: Segment Arb Export dialog box
3. Select the pulse card and channel for the waveform to be exported.
4. Select
...
.
5. Locate the target folder.