1-10
Return to
4200-900-01 Rev. K / February 2017
Section 1: Getting Started
Model 4200-SCS User’s Manual
Model 4200-SCS Hardware Overview
DC source-measure unit (SMU)
This section provides detailed information about several 4200-SCS hardware components, and is
arranged as follows:
•
Models 4200-SMU and 4210-SMU overview
: Discusses 4200-SMU and 4210-SMU basic
source and measure characteristics, basic circuit configurations, operating boundaries, and
connectors.
•
SMU with Model 4200-PA overview
Describes how the 4200-PA extends 4200-SMU and
4210-SMU dynamic range, and covers source and measure characteristics, basic circuit
configurations, operating boundaries, connectors, and mounting methods.
•
:
Provides basic information about using the ground unit, including
basic characteristics and connectors.
Models 4200-SMU and 4210-SMU overview
The following paragraphs discuss the basic characteristics of the 4200-SMU and 4210-SMU.
Basic SMU circuit configuration
The basic SMU circuit configuration is shown in
. The SMU operates as a voltage or
current source (depending on source function) in series with an I-Meter, and connected in parallel
with a V-Meter. The voltage limit (V-limit) and current limit (I-limit) circuits limit the voltage or
current to the programmed compliance value. In this local sensing example, the SMU FORCE
terminal is connected to device-under test (DUT) HI, while the DUT LO is connected to COMMON.
See the
Reference Manual, Connections and Configuration, page 4-1
Source-Measure Concepts, page 5-1
, for details.