Safety precautions 

Observe the following safety precautions before using this product and any associated 
instrumentation. Although some instruments and accessories would normally be used 
with nonhazardous voltages, there are situations where hazardous conditions may be 
present. 
This product is intended for use by qualified personnel who recognize shock hazards 
and are familiar with the safety precautions required to avoid possible injury. Read and 
follow all installation, operation, and maintenance information carefully before using the 
product. Refer to the user documentation for complete product specifications. 
If the product is used in a manner not specified, the protection provided by the product 
warranty may be impaired. 
The types of product users are: 

Responsible body

 is the individual or group responsible for use and maintenance of 

equipment, for ensuring that the equipment is operated within its specifications and 
operating limits, and for ensuring that operators are adequately trained. 

Operators

 use the product for its intended function. They must be trained in electrical 

safety procedures and proper use of the instrument. They must be protected from 
electric shock and contact with hazardous live circuits. 

Maintenance personnel

 perform routine procedures on the product to keep it 

operating properly, for example, setting the line voltage or replacing consumable 
materials. Maintenance procedures are described in the user documentation. The 
procedures explicitly state if the operator may perform them. Otherwise, they should be 
performed only by service personnel. 

Service personnel

 are trained to work on live circuits, perform safe installations, and 

repair products. Only properly trained service personnel may perform installation and 
service procedures. 
Keithley Instruments products are designed for use with electrical signals that are rated 
Measurement Category I and Measurement Category II, as described in the 
International Electrotechnical Commission (IEC) Standard IEC 60664. Most 
measurement, control, and data I/O signals are Measurement Category I and must not 
be directly connected to mains voltage or to voltage sources with high transient 
overvoltages. Measurement Category II connections require protection for high 
transient overvoltages often associated with local AC mains connections. Assume all 
measurement, control, and data I/O connections are for connection to Category I 
sources unless otherwise marked or described in the user documentation. Main supply 
voltage fluctations not to exceed ±10% of the nominal voltage. 
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present 
on cable connector jacks or test fixtures. The American National Standards Institute 
(ANSI) states that a shock hazard exists when voltage levels greater than 30 V RMS, 
42.4 V peak, or 60 V DC are present. A good safety practice is to expect that hazardous 
voltage is present in any unknown circuit before measuring. 
Operators of this product must be protected from electric shock at all times. The 
responsible body must ensure that operators are prevented access and/or insulated from 
every connection point. In some cases, connections must be exposed to potential 
human contact. Product operators in these circumstances must be trained to protect 
themselves from the risk of electric shock. If the circuit is capable of operating at or 
above 1000 V, no conductive part of the circuit may be exposed. 

 

 

Содержание 3700A Series

Страница 1: ...Series 3700A System Switch Multimeter Quick Start Guide Series 3700A System Switch Multimeter Quick Start Guide...

Страница 2: ...work on live circuits perform safe installations and repair products Only properly trained service personnel may perform installation and service procedures Keithley Instruments products are designed...

Страница 3: ...al levels of the instruments and accessories as defined in the specifications and operating information and as shown on the instrument or test fixture panels or switching card When fuses are used in a...

Страница 4: ...transformer test leads and input jacks must be purchased from Keithley Instruments Standard fuses with applicable national safety approvals may be used if the rating and type are the same Other compon...

Страница 5: ...system switch with high performance DMM 3706A NFP six slot system switch high performance DMM without front panel display and keypad 3706A S six slot system switch without DMM 3706A SNFP six slot sys...

Страница 6: ...n including PDFs of this Quick Start Guide User s Manual Reference Manual Switching and Control Cards Reference Manual the product data sheet product specifications and rack mount kit instructions IVI...

Страница 7: ...for damage 2 Open the top of the box 3 Remove the packaging insert Packaging insert 4 Remove the bag that contains the documentation accessories and CD ROMs Packaging insert Series 3700A documentatio...

Страница 8: ...lowing Series 3700A System Switch Multimeters Model 3706A six slot system switch with high performance DMM Model 3706A S six slot system switch without DMM Model 3706A NFP six slot system switch with...

Страница 9: ...1 Series 3700A Product Information and Test Script Builder CD ROMs 2 Power line cord 3 RJ 45 LAN crossover cable 4 Safety Standards Conformance Information 5 Series 3700A Quick Start Guide Refer to t...

Страница 10: ...tiplexer Card Model 3723 Dual 1 30 High Speed Reed Relay Multiplexer Card To keep users safe always read and follow all safety warnings provided with each of the instruments in your system Model 3724...

Страница 11: ...card is already connected to a device under test DUT make sure power is removed from all external circuitry NOTE For inexperienced users disconnect DUT and external circuitry from switching cards Thi...

Страница 12: ...align the card edge into the card guide of the slot Item Description 1 Card guide part of Series 3700A instrument 2 Card 3 Card edge part of card 4 Mounting screw part of card 6 Slide in the card For...

Страница 13: ...the side panel of the Series 3700A instrument The space around these areas should be free from obstruction to ensure proper fan operation Connect line power Series 3700A instruments operate from a lin...

Страница 14: ...ition 2 Connect the socket of the supplied power cord to the power connection on the rear panel Connect power cord 3 Connect the plug of the power cord to a grounded AC outlet Turn on the instrument T...

Страница 15: ...t panel when the instrument is turned on you should see V V AUTO AUTO DCV R AUTO N 1 0000 AD DCV R AUTO N 1 0000 AD When initialization is complete you will see the default display screen as shown bel...

Страница 16: ...ystem information menu 4 Turn the navigation wheel to select SERIAL 5 Press the navigation wheel to view the serial number 6 Press EXIT several times to return to the main display These steps confirm...

Страница 17: ...r s Manual Next steps Refer to the Series 3700A User s Manual This manual provides application based examples that will help familiarize you with the instrument This manual is on the Product Informati...

Отзывы: