Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-177
The filter type setting is saved with the
dmm.func
function setting, so if you use another function,
then return to the previous function, the filter type setting you set previously is retained.
Example
dmm.func = "twowireohms"
dmm.filter.type = dmm.FILTER_MOVING_AVG
dmm.filter.enable = dmm.ON
Set the filter type for 2-wire ohms to moving
average and enable filtered measurements.
Also see
(on page 11-166)
(on page 11-174)
(on page 11-175)
(on page 11-177)
(on page 11-179)
dmm.filter.window
This attribute sets the window for the averaging filter that is used for measurements for the selected function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Reset
DMM reset
Recall setup
Create configuration script
Save setup
1.000000E-01 (0.1)
Usage
value
= dmm.filter.window
dmm.filter.window =
value
value
The filter window setting; the range is between 0 and 10 to indicate percent of range
Details
The filter window setting only applies when
dmm.func
is set to one of the following:
•
"accurrent"
•
"acvolts"
•
"commonsideohms"
•
"dccurrent"
•
"dcvolts"
•
"fourwireohms"
•
"temperature"
•
"twowireohms"
Querying the setting for any other function returns
nil
. An error is generated if this setting is written
or read for any other function.
An error is generated if the value is out of range.
The filter window setting is saved with the
dmm.func
function setting, so if you use another function,
then return to the previous function, the filter window setting you set previously is retained.