Series 3700A System Switch/Multimeter Reference Manual
Section 11: TSP command reference
3700AS-901-01 Rev. D/June 2018
11-157
Example
dmm.calibration.save()
dmm.calibration.lock()
Save calibration, then lock it.
Also see
(on page 11-158)
(on page 11-157)
dmm.calibration.password
This attribute sets the password that must be entered before you can unlock calibration.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (W)
Yes
Not applicable
DMM nonvolatile memory KI003706
Usage
dmm.calibration.password = "
password
"
password
A string that represents the valid password to unlock calibration
Details
This attribute can only be set when calibration is unlocked.
This attribute generates an error if calibration is locked or if the password string length is greater than
ten characters.
Be sure to record the password; there is no command to query for the password once it is set.
Example
dmm.calibration.unlock("KI003706")
dmm.calibration.password = "myUnlock"
dmm.calibration.lock()
To change the default calibration password,
unlock the calibration with the default
password.
Saves the password as
myUnlock
.
Lock calibration. Subsequent unlocks will use
the password
myUnlock
.
Also see
(on page 11-158)
dmm.calibration.save()
This function saves calibration data.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
dmm.calibration.save()